User's Guide _______________________________________________________________________
60 ___________________________________________________________________ M211399EN-G
Clear User Adjustment of T Measurement
CT
[
RESET
]<cr>
Example:
>
pass 9000
>
ct reset
OK
Enter Calibration and Adjustment
Information
Use the
CTEXT
command to store a text string that describes the
calibration and/or adjustment. To enter a text string with spaces, enclose
the string in quotation marks. Use the
CDATE
to store the date.
CTEXT
[
text
]<cr>
CDATE
[
YYYY-MM-DD
]<cr>
Examples:
>
pass 9000
>
ctext “adjusted rhlab/Tech021”
“adjusted rhlab/Tech021”
>
cdate 2011-12-08
Calibration date : 2011
-
12
-
08