6.
Press
OK
to start adjustment.
7.
Check the environment settings if needed. Otherwise press
NO
.
Continue according to the directions of the desired adjustment
method.
Field Checking and Adjustment Using a Calibrated
Reference Probe
Follow steps 1 to 7 on page 25 and continue as follows:
8.
Check that the probes are located in equal conditions and wait
until the readings have stabilized (may take 30 minutes or
more). If you are near the probes, do not breath in their
direction.
9.
Press
ADJUST
to continue adjusting.
10.
Choose
To same as RH
I/II
from the MI70 adjustment menu, press
SELECT
(MI70 automatically recognizes which port the HMP70-
series probe is connected to).
11.
Confirm the adjustment by pressing
YES
.
12.
The adjustment is done. Press
BACK
and
EXIT
to return to the
basic display.
13.
Turn off the MI70 and detach the connection cable.
One-Point Adjustment Using a Calibrator
When adjusting the transmitter in only one reference condition, please
take care that the reference condition represents the measuring
environment. MI70 indicator is used now only as a terminal for
visualizing and setting the transmitter's RH reading.
Follow steps 1 to 7 on page 25 and continue as follows:
8.
Remove the filter from the transmitter's probe and insert the
probe head into the reference condition.
9.
Press
ADJUST
to continue adjusting.
10.
Choose
1-point adjustment
from the MI70 adjustment menu, press
SELECT
.