Register
number
Register
address
Register description
Data format
Unit / Valid range
General
2561
0A00
hex
User information
Text
Text string of 24 bytes
in UTF-8 encoding
0A01
hex
0A02
hex
0A03
hex
0A04
hex
0A05
hex
0A06
hex
0A07
hex
0A08
hex
0A09
hex
0A0A
hex
0A0B
hex
Compensation setpoints
769
0300
hex
Pressure compensation
setpoint
32-bit float
Unit: hPA
Default: 1013.25 hPa
0301
hex
821
0334
hex
Temperature compensation
setpoint. If a value is written
to this register, the device
uses it instead of its own
temperature measurement.
When probe heating is in use,
temperature must be written
to this register to enable
output parameters that
depend on temperature
measurement (for example,
relative humidity).
32-bit float
Unit: °C
0335
hex
Purge
773
0304
hex
Purge interval
32-bit float
Unit: min
10 ... 14400
0305
hex
1283
0502
hex
Interval purge on/off
16-bit boolean
0 = Off
1 = On
HMM170 User Guide
M212259EN-D
38