A.4.1 Measurement Data Registers
Table 13 Modbus Measurement Data Registers (Read-Only)
Address
(Decimal)
Address
(Hexadecimal)
Register Description
Data
Format
Unit
1
0000
hex
Vaporized hydrogen peroxide
concentration by volume
32-bit float ppm
3
0002
hex
H
2
O+H
2
O
2
relative saturation
32-bit float %RS
5
0004
hex
Temperature used for calculation. By
default, shows the temperature from
the temperature probe (register
0016
hex
).
If temperature compensation is turned
ON (register 0505
hex
), this register
shows the volatile value for given
temperature reading (register 0302
hex
).
See
.
32-bit float °C
7
0006
hex
Relative humidity
32-bit float %RH
9
0008
hex
Absolute hydrogen peroxide
32-bit float mg/m
3
15
000E
hex
Water concentration by volume
32-bit float ppm
17
0010
hex
Water vapor pressure
32-bit float hPa
19
0012
hex
Absolute humidity (water)
32-bit float g/m
3
21
0014
hex
Water vapor saturation pressure
32-bit float hPa
23
0016
hex
Temperature from the temperature
probe
32-bit float °C
257
0100
hex
Vaporized hydrogen peroxide
concentration by volume
16-bit
signed
integer
ppm
258
0101
hex
H
2
O+H
2
O
2
relative saturation
16-bit
signed
integer
%RS * 100
259
0102
hex
Temperature used for calculation. By
default, shows the temperature from
the temperature probe (register
010B
hex
).
If temperature compensation is turned
ON (register 0505
hex
), this register
shows the volatile value for given
temperature reading (register 0302
hex
).
See
.
16-bit
signed
integer
°C * 100
Appendix A – Modbus Reference
47