Table 14 Modbus Configuration Data Registers (Writable)
Address
(Decimal)
Address
(Hexadecimal)
Register Description
Data
Format
Unit / Valid Range
769
0300
hex
Volatile value for
pressure compensation
(value cleared at probe
reset). Used for
compensating the
measurement for
pressure if pressure
compensation is turned
ON (register 0504
hex
).
32-bit float hPa
813.25 ... 1213.25
(default 1013.25 hPa)
(Init copied from power-
up value)
771
0302
hex
Volatile value for given
temperature reading
(value cleared at probe
reset). Used for
calculations instead of
the temperature from the
temperature probe if
temperature
compensation is turned
ON (register 0505
hex
).
32-bit float °C
-40 ... +80
(Init copied from power-
up value)
773
0304
hex
Power-up value for
pressure compensation
32-bit float hPa
813.25 ... 1213.25
(default 1013.25 hPa)
775
0306
hex
Power-up value for given
temperature reading
32-bit float °C
-40 ... +80
777
0308
hex
Purge interval
16-bit
integer
min
60 ... 10080
(= 1 hour ... 1 week)
(default: 1440 (= 24
hours))
Appendix A – Modbus Reference
49