4.1 Wiring
1
5
3
4
2
Figure 7 Probe M12/5 Pins
Pin #
Function
Notes
Wire
Color
1
Power supply
With digital output: 15 ... 30 VDC
With analog output: 15 ... 25 VDC
When using analog outputs, it is recommended to
use a low supply voltage to minimize self-heating
and maximize measurement performance.
Brown
2
RS-485- or analog
output 2
Current output: 4 … 20 mA (default).
White
3
Power and signal GND
Blue
4
RS-485+ or analog
output 1
Current output: 4…20 mA (default)
Black
5
Output control and purge
trigger in analog mode
Floating = RS-485
Grounded = Analog outputs
If you want to be able to trigger a purge manually in
the analog mode, do not connect pin #5
permanently to ground, but instead, use a relay or
similar to control the pin.
Grey
1)
Wire colors apply to the following cables: 223263SP, 26719SP, 26720SP, 216546SP,
244669SP
2) The ordered parameters and scaling are shown in the calibration certificate delivered with
the probe.
HPP272 User Guide
M211972EN-B
22