The probe head can be mounted through thin metal walls using the included cable gland or
mounting grommet. Two grommets are included: small one for 6.5 mm diameter hole, and
large one for 12.5 mm diameter hole.
You can also attach the probe head directly using a zip tie. The probe head should be attached
from the point near the black plastic part.
• Temperature measurement range −40 … +120 °C (−40 … +248 °F)
• Operating temperature of probe body −40 … +60 °C (−40 … +140 °F)
• Integrated filter (non-replaceable)
Do not damage the probe head by bending, crushing, or striking it.
Avoid overtightening when installing the probe head through a cable gland.
CAUTION!
3.6.1 Installing HMP9 through a cable gland
1
2
3
4
Figure 10 Installing HMP9 probe head through a cable gland
1
Black plastic part of the HMP9 probe head
2
Nut for tightening the probe in place
3
Base of the cable gland
4
M10×1.5 threads of the cable gland
• M10×1.5 cable gland (included with HMP9 probe)
• Drill with 8.5 mm bit
• M10×1.5 threading tap
• 13 mm wrench
1. Drill a 8.5 mm diameter hole in the installation location.
HMP3, HMP4, HMP5, HMP7, HMP8, HMP9, MMP8, TMP1 User Guide
M212022EN-D
22