Error
Likely Cause
Suggested Solution
Capacitance reference
error. [47]
Internal transmitter
failure.
Power-cycle the device, and if necessary, restore
the factory settings. If the error remains, contact
a Vaisala Service Center.
Firmware checksum
mismatch. [49]
Device settings
corrupted. [50]
Additional
configuration settings
corrupted. [51]
Sensor coefficients
corrupted. [52]
Main configuration
settings corrupted.
[53]
Non-volatile memory
read/write failure. [57]
Calibration certificate
checksum mismatch.
[58]
In case of constant error, contact a Vaisala Service Center.
More Information
‣
8.3 Error Codes in MI70 Handheld Meter
If several errors are active at the same time, the MI70 shows the sum of currently active error codes.
The label "
hex
" might not be displayed in MI70.
Refer to
Error Messages in Insight Software (page 70)
for more information about the error states.
Table 43 Error Codes in MI70
Transmitter Error Code in MI70
Corresponding Error
1
hex
Temperature measurement error. [44]
2
hex
Humidity measurement error. [45]
4
hex
Humidity sensor failure. [46]
8
hex
Capacitance reference error. [47]
Chapter 8 – Troubleshooting
71