TECHNICAL DATA

Carbon dioxide

Measuring range

0...2000 ppm CO

2

 (nominal); with recalibration 0...5000 ppm CO

2

,

0...10 000 ppm CO

2

, 0...20 000 ppm CO

2

Accuracy at 20°C against
certified factory references

<

±[

30 ppm + 2.0 % of the reading

]

(including repeatability and calibration uncertainty)
Non-linearity

<

±

1.0 %FS

Temperature dependence
of output (typical)  

< 0.15 %FS / °C (reference 25

°

C)

Long-term stability 

<

±

5 %FS / 5 years

Response time (0…63%)

1 minute

Operating conditions

Operating temperature

-5...+45 °C

Storage temperature

-20...+70 °C

Humidity range

short term 0...85 %RH  (non-condensing)

Air flow range

0...10 m/s

General

Analogue outputs

0...20 mA or 4...20 mA and  0...10 V

Resolution of analog outputs  0.5 % FS
Recommended external load:

current output

max. 500 

         

voltage output       

min. 1k

Relay contact ratings

max. 50 V 0.5 A

Power supply

nominal 24 VDC/VAC  (18...30 VDC)

Power consumption

< 2.5 W average

Warm-up time

< 15 minutes

Mechanics

Housing material

ABS plastic

Housing classification

IP65

(electronics housing)
Weight:
GMD20

140 g

GMD20D

170 g

ACCESSORIES

Order code

Description

GMR20

Relay output option (not for the GMD20D)

GML20

LonWorks

interface option (not for the GMD20D )

GM35001

Calibration pipe

19222GM

Calibration software kit (incl. disk and serial COM adapter)

GUARANTEE

Vaisala issues a guarantee for the material and workmanship of this product under
normal operating conditions for one (1) year from the date of delivery. Exceptional
operating conditions, damage due to careless handling and misapplication will void
the guarantee.

              

CO

2

 

transmitter GMD20 and GMD20D

                 

M210292en-A

4

*M

210292

EN

*

Reviews: