USER'S GUIDE____________________________________________________________________
90 __________________________________________________________________ M210762EN-D
To select display units:
1. Open
the
MAIN MENU
by pressing any of the
▲▼◄►
arrow
buttons.
2. Press
the
►
arrow button to select
Display
.
3. Use
the
▲▼
arrow buttons to select
Units
. Confirm the selection
by pressing the right-hand arrow button.
4. Use
the
▲▼
arrow buttons to select display units. Confirm the
selection by pressing
CHANGE
. The unit changes from metric to
non-metric or the other way round.
5. Press
EXIT
to return to the basic display.
NOTE
Changing the units by using the display/keypad has no effect on the
serial output units.
Using Serial Line
Use the serial line command
FORM
to change the format, and the
command
UNIT
to select metric or non-metric output units.
FORM
Use the serial line command
FORM
to change the format or select a
certain quantities for the output commands
SEND
and
R
.
Syntax:
FORM
[
x
]
where
x = Formatter
string
Formatter string consists of quantities and modifiers. If no formatter
string is entered, the command will display the currently active formatter
string. Note that the hash symbol "#" is shown as a backslash "\" when
viewing the current formatter string.
When selecting the quantity, use the abbreviations of the quantities
presented in Table 5 on page 20 when issuing the command. The
modifiers are presented in Table 26 below.
Table 26
Modifiers
Modifier Description
x.y
Length modifier (number of digits and decimal places)
#t Tabulator
#r Carriage-return
#n Line
feed