_________________________________________________________________________ 65
•
the LED on the left flashes; set the low end of channel 2 with the
arrow keys and press ENT
•
the second LED from the left flashes; set the high end of channel 2
with the arrow keys and press ENT
The analogue outputs are calibrated to ensure that outputs are
correctly scaled: for example, when the output is scaled to 4...20 mA,
the low end of the scale is 4 mA and high end 20 mA exactly.
However, when 0... 20 mA output is used, the output can not be
adjusted to exactly 0 mA, but to 50 µA. When 0...1 V, 0...5 V or 0...10
V output is in use, the output is adjusted to 50 mV. The following
table summarizes the correct output values.
T
ABLE
6-2
Summary of the correct output values
Output scale:
0...20 mA
4...20 mA
0...1 V
0...5 V
0...10 V
low end:
50 µA
4 mA
50 mV
50 mV
50 mV
high end:
20 mA
20 mA
1 V
5 V
10 V
All manuals and user guides at all-guides.com
Summary of Contents for DMP248
Page 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 28: ...All manuals and user guides at all guides com...
Page 36: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 70: ...All manuals and user guides at all guides com...
Page 82: ...All manuals and user guides at all guides com...
Page 84: ...All manuals and user guides at all guides com...
Page 108: ...All manuals and user guides at all guides com...
Page 133: ...APPENDIX 6 All manuals and user guides at all guides com...
Page 135: ...All manuals and user guides at all guides com...
Page 140: ...All manuals and user guides at all guides com...