Chapter 8 ___________________________________________________________________ Repair
VAISALA ______________________________________________________________________ 103
WARNING
Disconnect the CT3839 power cable before continuing!
3.
Remove the no-break battery. Refer to steps 3 to 5 in section
Replacing No-break Battery on page 101.
4.
Remove the entire battery cage by removing the three screws
holding it in its place. One of the screws is located underneath
the battery cage, the other two are on the left side near the
bottom of the cage.
5.
Detach the data line connector.
6.
Remove the six attachment screws from the bottom of the
ceilometer measurement unit. Also disconnect the AC power
cable. Lift the AC power unit from the ceilometer measurement
unit.
7.
Put the new AC power unit in place of the old one and connect
the AC power cable. Tighten the six attachment screws back to
their places.
8.
Attach the data line connector again.
9.
Put the no-break battery cage back to its place and tighten the
three screws.
10.
Replace the no-break battery as you removed it. Refer to steps 6
to 8 in section Replacing No-break Battery on page 101.
11.
Connect the CT3839 power cable.
12.
Switch the power on with all three switches. Wait until the
Laser on
LED starts blinking at 2-second intervals. Ensure that
all six diagnostic LEDs light up after the self-test. Refer to
Figure 19 on page 99.
The ceilometer unit should now return to normal operation. Verify
that the unit is working properly and that there are no other failures
with the system. If other failures exist, separate troubleshooting may
be necessary.