Figure 45
Error Indicator and Error Message
The error indicator disappears when the error state is over and you have
checked the error message. Press the
INFO
button to display the error
message.
You can also check the error message through the serial interface by
using the command
ERRS
. In case of constant error, please contact
Vaisala, see
Vaisala Service Centers on page 125
The number refers to
1
=
Error Indicator
Table 14
Error Messages
Error Code Error Message
Action
E10
Internal EEPROM read error.
Internal barometer failure. Return the barometer
to the Vaisala Service Center.
E11
Internal EEPROM write error.
Internal barometer failure. Remove the barometer
and return the faulty unit to Vaisala Service.
E12...E15
Add-on module 1/2/3/4 connection
failure
Turn off the power and check the module
connection. Turn on the power.
E8
Device internal temperature out of
range
Ensure that the operating temperature is within
the valid range
E6
Operating voltage out of range
Ensure that the operating voltage is within the
valid range.
E7
Internal system voltage out of
range
Internal barometer failure. Return the barometer
to the Vaisala Service Center.
E20...E23
Configuration switches for analog
output 1/2/3/4 set incorrectly
Check and re-set the switches of the analog
output module, see section
E5
Communication module installed
in incorrect add-on module slot
Disconnect the power and change the
communication module to module slot 1.