User's Guide ______________________________________________________________________
230 _________________________________________________________________ M211296EN-A
Command Connection to WMT703
In order to troubleshoot or reconfigure WMT703, you may need to open
a command connection to the sensor as follows:
1.
Start AWS Client software.
2.
From the AWS Client menu, select
Settings
-
Options
.
3.
Create two new function key mappings as shown in Figure 123
below.
- $0OPEN<CR><LF>
- <CR><LF>
1101-040
Figure 123
Defining Function Key Mappings for WMT703
4.
Connect to QML201 logger as described in section Opening
Service Connection on page 148.
5.
Open command connection on QML201 by typing
open<CR>
to
the AWS Client terminal window.
6. Type
command
open COM1<CR>.
This should output a response
Terminal I/O re-routed to COM1
.
7.
Press F11 (or other function key where the
open
command was
programmed to in step 3 above) to send the
open
command to
WMT703.