6. To change the calibration values, use the command line interface.
Use the following parameters to adjust internal gas measurements:
•
co_span
•
no2_span
•
o3_span
•
so2_span
•
co_zero
•
no2_zero
•
o3_zero
•
so2_zero
The span adjustment is proportional and is represented as a percentage value, which
can vary between
1
...
255
(default =
100
). The zero adjustment is an absolute value in
ppb and it is added after span adjustment. The zero values can vary between
-10000
...
10000
. Use the
show
command to read the concurrent values before using
the
set
command to replace them. If AQT400 has an offset adjustment (zero value
other than
0
), take the old value in count before setting a new one. This also applies to
span values.
7. Use the
write
--really
command to save the new adjustments.
8. Use the
show
health
command to check the health status of AQT400. The output is a
percentage value (0 ... 100 %). If the health value is below
20
, contact Vaisala technical
support.
Appendix A – Performance Check
41
Summary of Contents for AQT400
Page 1: ...M211930EN A User Guide Air Quality Transmitter AQT400...
Page 4: ...Index 45 Technical Support 47 Warranty 47 Recycling 47 AQT400 User Guide M211930EN A 2...
Page 8: ...Indicates that you need to take some notes during the task AQT400 User Guide M211930EN A 6...
Page 12: ...AQT400 User Guide M211930EN A 10...
Page 14: ...AQT400 User Guide M211930EN A 12...
Page 18: ...AQT400 User Guide M211930EN A 16...
Page 20: ...AQT400 User Guide M211930EN A 18...
Page 26: ...AQT400 User Guide M211930EN A 24...
Page 30: ...AQT400 User Guide M211930EN A 28...
Page 32: ...AQT400 User Guide M211930EN A 30...
Page 34: ...AQT400 User Guide M211930EN A 32...
Page 40: ...AQT400 User Guide M211930EN A 38...
Page 44: ...AQT400 User Guide M211930EN A 42...
Page 46: ...AQT400 User Guide M211930EN A 44...
Page 48: ...AQT400 User Guide M211930EN A 46...
Page 50: ...AQT400 User Guide M211930EN A 48...
Page 51: ......
Page 52: ...www vaisala com...