Cleaning and maintenance
7.4
7.4 Replacing the device fuses
Replacing the device fuses
Vacuum pumps with switchable dual‑voltage motor have two
device fuses. Vacuum pumps in 230 V – version have no device
fuse. The device fuses are located laterally at the terminal box,
Type: 250 VAC / 10 AT – 5x20.
DANGER
DANGER
Danger due to electrical voltage.
The vacuum pump has two fuses (double pole / neutral
fusing) and might be energized, even when one fuse
has blown or has been removed.
>
Switch off the vacuum pump.
>
Unplug the power plug from the socket, before
unscrewing the fuse holder.
Replacing the device fuse
1.
1. Identify and eliminate the cause of failure before switching on
the pump again.
2.
2.
Unscrew the fuse holder at the terminal box of the pump
using a slotted screwdriver.
3.
3. Replace the defective fuse by a fuse of the same type
see
chapter:
.
4.
4.
Reassemble the holder with fuse to the pump; slotted screw
‑
driver.
Replacing the device
fuse