page 30 of 121
Gas inlet temperatures
Wetted parts
Components
Wetted materials
Pump
Housing cover
PTFE
Head cover
ETFE carbon fiber reinforced
Diaphragm clamping disc
ETFE carbon fiber reinforced
Diaphragm
PTFE
Valve
FFKM
Pumping unit
Inlet pumping unit
PPS (PC 3001 VARIO
pro
IK: PP)
Outlet pumping unit
PET (PC 3001 without emission con-
denser: PTFE carbon reinforced)
Outlet pumping unit (PC 3001 VARIO
pro
emission condenser Peltronic)
ETFE / ECTFE
Tubing
PTFE
Screw-in fittings
ETFE / ECTFE
O-ring at the catchpot
Fluoroelastomer
Overpressure safety relief device
at exhaust waste vapor condenser
Silicone rubber / PTFE film
Distribution head (inlet)
PPS glass fiber reinforced /
PP (blind plug)
Waste vapor condenser IK / EK / TE,
catchpot
Borosilicate glass
Emission condenser Peltronic
ETFE / ECTFE / PP / PA
Silencer
PBT / PVF / rubber
CVC 3000
Sensor
Aluminum oxide ceramic
Sensor housing
PPS / glass fiber
Sensor seal
Chemically resistant fluoroelastomer
Venting valve seal
FFKM
We reserve the right for technical modification without prior notice!
* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)
Operating condition
Inlet pressure
Permitted range of gas
temperatures at inlet
Continuous operation
> 75 Torr (100 mbar)
(high gas load)
➨
50 °F to 104 °F
(+10°C to +40°C)
Continuous operation
< 75 Torr (100 mbar)
(low gas load)
➨
32 °F to 140 °F*
(0°C to +60°C*)
Short-time
(< 5 minutes)
< 75 Torr (100 mbar)
(low gas load)
➨
14 °F to 176 °F*