68
20901115_12+16C_VARIO_sel_V1.0_130618
Cleaning and maintenance
10
11
10.
Place both valve terminals
with the disc springs on the
pump heads.
11.
Place the clamping claws
on the valve terminals and
tighten the screw fittings first
by hand, then with a torque
wrench to 3 Nm.
12
13
12.
Carefully press the dia‑
phragms centrally into the
housing opening, ensuring
they are flush with it.
13.
Hold the pump head pair at
the vacuum pump and wind
in the screw fittings; hex key
size 5.
14
15
14.
Slide the molded hoses back
onto the hose nozzles.
15.
Secure the hose clips on the
hose nozzles, e. g., with flat
nose pliers.
Summary of Contents for MV 10C VARIO select
Page 26: ...26 20901115_EN_Mx10 12 16C_VARIO_sel_V1 0_130618 Product description...
Page 38: ...38 20901115_EN_Mx10 12 16C_VARIO_sel_V1 0_130618 Installation and connection...
Page 86: ...86 20901115_EN_Mx10 12 16C_VARIO_sel_V1 0_130618...
Page 87: ...87 20901115_EN_Mx10 12 16C_VARIO_sel_V1 0_130618...