Establish hose connections to the pump head in three
places.
➨
Push transparent PVC hoses (2x) onto the hose connec-
tors (1)+(2).
➨
Push PTFE tubing onto the connector at the quick cou-
pling (3). Fix the PTFE tubing with the clip to the hose
connector. Close clip with flat pliers.
Assembling the connecting hoses
Reassemble in reverse order.
➨
Install head cover, valves and housing cover with housing
cover insert.
☞
Make sure that the
valves are correctly seated
: Valves at
the outlet with round centred opening under valve, valves
at the inlet with kidney-shaped opening beside valve.
➨
By turning eccentric bushing, bring connecting rod into up-
per turning point position (Maximum stroke of the rod).
☞
Pay attention that the diaphragm stays positioned centrally
so that it will become clamped uniformly between housing
and head cover.
➨
Screw in four socket head screws fixing housing cover
crosswise first slightly then tighten.
☞
Do not tighten until head cover is in contact with housing,
torque
12 Nm
.
Assembling the pump head
➨
By turning eccentric bushing (front of connecting rod),
bring connecting rod into a position in which diaphragm is
in contact with housing and centred with respect to bore.
☞
Tilt pump carefully onto the left side so that the diaphragm
is on top. Support pump appropriately.
1
3
2