UWE SEK-12K Operation Manual Download Page 8

Page: 14

INTERNAL FUNCTIONS

c) After Baud rate is selected, press MODE/RANGE, and the

display will show “P=XXX”. Select Parity Setting by
pressing SAMPLE.

P=n81

no Parity bit, data = 8 bit, stop bit = 1 bit

P=E71

Even Parity bit, data = 7 bit, stop bit = 1 bit

d) Press TARE to confirm.
e) Enter function F7 to select RS-232 transmission mode
f) Start RS-232 software on the PC.
g) When placing an item on the platter, after weight becomes

stable, the computer display will show a line of RS-232
data.

S

W W W .

W W

U

U

CR LF

Blank

Blank

eg

.

-

3

0

0

0

.

1

g

CR LF

6

1

3

.

5

1

b

CR LF

(F6) Enable/Disable Digital Filter

Digital Filter can be enabled when weighing in environment

with external interferences, such as wind, vibration or other
unknown interference. The scale will have a more stable reading;
however, the speed response will be decreased.

Page: 15

INTERNAL FUNCTIONS

a) Use the procedure described above to reach Function F6.
b) Press MODE/RANGE and display will show “FiLt_X” (X

being the number selected from previously)

c) Press MODE/RANGE to select the Digital Filter number

required.

FiLt_0

Disable Digital Filter

FiLt_1~5

Enable Digital Filter.

(Higher number represents a stronger filter with more weighing
stability but with slower weighing response speed)

d)

Press TARE to confirm, then press ON/ZERO to return to
normal weighing mode.

(F7) RS-232 Transmission Mode

a) Use the procedure described above to reach Function F7.
b) Press MODE/RANGE, and the display will show “XXXX”

(XXXX being the setting from the previous selection)

c) Select RS-232 Transmission mode by pressing MODE.

Auto

Send data once after weight stabilization

P-Out

Printer Output (**Reserved)

Conti

Send data continuously after weight
stabilization

d) Press TARE to confirm, then press ON/ZERO to return to

normal operation.

Summary of Contents for SEK-12K

Page 1: ...ORE OPERATING THE SCALE Specifications and functions subject to change without prior notice X0SEK030000 Revision April 2008 Page 1 CONTENT 1 CAUTION 2 2 SPECIFICATIONS 3 3 FEATURES 4 4 KEYBOARD LAYOUT FUNCTIONS 5 5 DISPLAY INDICATORS 7 6 INTERNAL FUNCTIONS 8 7 TROUBLESHOOTING 14 8 CALIBRATION 15 ...

Page 2: ...ial acetone volatile solvent thinner or alcohol for cleaning g Do not use pressurized water nozzle for cleaning h The scale must be placed horizontally during transportation or long time storage i Remove platter from scale before transportation or long time storage j Store scale in a dry and clean place Page 3 SPECIFICATIONS Model No Capacity Max Division e SEK 3000 3000g 6 6lb 0 05g 0 01oz SEK 60...

Page 3: ...rs for settings for example setting the HI OK LO 2 0 STANDARD When in regular use press this key to input 0 When inputting HI OK LO value or a load is placed on the scale press and hold this key to set the weight displayed as the OK value The LED for OK will light up when the value is stored in the scale 3 SAMPLE When the counting pcs function is in use press this key for the sampling function Use...

Page 4: ...0 1 g 6000 2 g 6000 5 g 0 0000 lb 0 0 00 lb lb oz 0 0 0 0 Tael PCS 6000 1 g Page 7 KEYBOARD LAYOUT FUNCTIONS When inputting HI OK LO value or a load is placed on the scale press and hold MODE RANGE to set the displayed value as the range for HI and LO values from the OK value The LED for HI and LO will light when the value is stored in the scale Remark The weighing units can be turned on off in th...

Page 5: ...rated capacity To maintain normal operation please remove all weight from the platter before turning on the scale Page 9 DISPLAY INDICATORS 13 CHARGE STATUS LIGHT This LED light displays the charging status When charging using the adaptor the LED will turn red The LED will turn green once the rechargeable battery is fully charged 14 LED for LO This LED lights up when the weight or pcs on the scale...

Page 6: ...N ZERO 2 Display will show F0 3 Press TARE until the desired Function is displayed 4 Press MODE RANGE to enter function 5 Press MODE RANGE to make selection 6 Press TARE to confirm selection 7 Press ON ZERO key to return to weighing mode F0 Set the Auto Zero Tracking range Zero_X can select how many divisions of e to Auto Zero Within the setting range the weight is forced to be zero F1 Span Value ...

Page 7: ... units f In order to set the default weighing unit shown when the scale is first turned on press SAMPLE until the desired weighing unit is shown then press ON ZERO The display will show InIt 0 X g F4 Turn off and then turn on the scale to complete this process Page 13 INTERNAL FUNCTIONS F4 Auto Power off Function a Use the procedure described above to reach Function F4 b Press MODE RANGE The displ...

Page 8: ... the speed response will be decreased Page 15 INTERNAL FUNCTIONS a Use the procedure described above to reach Function F6 b Press MODE RANGE and display will show FiLt_X X being the number selected from previously c Press MODE RANGE to select the Digital Filter number required FiLt_0 Disable Digital Filter FiLt_1 5 Enable Digital Filter Higher number represents a stronger filter with more weighing...

Page 9: ...g the setting from the last selection nobEEP disable sound alarm gobEEP sound alarm when OK value is reached HLbEEP sound alarm when HI LO value is reached HbEEP sound alarm when HI value is reached LbEEP sound alarm when LO value is reached c Press MODE RANGE to select the desired setting Page 17 INTERNAL FUNCTIONS d Press TARE to confirm then press ON ZERO to return to normal weighing mode F10 R...

Page 10: ...rference 1 Remove the object and try again 2 Go to an environment free of interference and try again Page 19 CALIBRATION 1 Press ON ZERO to turn on the scale Before the countdown finish from 9 to 0 press and hold MODE RANGE key 2 The display will show CAL _1 3 Press MODE RANGE 4 The display will show YES The platter should be free from any loads to calibrate the zero point 5 After zero point is ca...

Reviews: