12. The user use 3M8514 welding protection mask (or equal grade protection masks) when use ozone to 

disinfect, and can process ozone protection which at 10 times occupation touch limit potency under 
1PPM

13. This product set enough safety device to guarantee the safety of users. The daily item to item check is 

the necessary condition to guarantee that safety use the equipment. Please process item to item check 
according to the requirements in the specification, need stop use the equipment immediately once 
found the problem. 

14. Because of the instability of ozone, ultraviolet light accelerates the decomposition of ozone and 

converts it back to oxygen. After the ozone work is completed, the equipment switches to UV mode to 
quickly reduce the ozone concentration in the equipment. The user must never open the cabinet doors 
before the device has completed the ozone disinfection operation (before the end of the timer 
countdown in the monitor)

15. Users are advised to equip users with ozone concentration detection instruments to regularly detect 

whether the leaked ozone value of the equipment exceeds the safe value: long term populated 
enclosed spaces, such as houses, apartments, hospitals and offices, ozone must be less than 0.1PPM 
in English. If ozone leakage value is found to have exceeded the above standard, the equipment must 
be immediately stopped.

To guarantee the disinfect effect of UVC:

1.   Whether UV light can effective disinfect depends on the amount of exposure received by the 

disinfection object, which depends on the radiation power of the UV light source, the distance of the 
disinfection object from the UV light source, and the duration of the exposure to UV light.

2.  When killing common microorganisms, the irradiated dose should reach 10,000 μW/cm²; 100,000 μ

W/cm² is the recommended dose to kill bacterial spores. The radiation dose should not be less than 
100,000 μW/cm² when the target microorganism for disinfection is unknown. 

3.  According to the radiation intensity of the UV light source of this product and the general distance of 

the UV light source from the irradiated object, it is recommended that:

•    In order to achieve the radiation dose of 10,000 μW/cm² to meet the general needs of daily use, the 

exposure time should be at least above 60 seconds

•    In order to achieve the radiation dose of 100,000 μW/cm² to meet the needs of the professional 

environment, the exposure time should be at least above 500 seconds

To guarantee the ozone disinfect effect:

1.   Ozone can kill bacterial reproductions, viruses, fungi, etc., and can destroy botulinum toxin. In order for 

ozone to be effective as a disinfectant, its concentration must be much higher than that that can be 
safely tolerated by humans and animals. Ozone has a powerful killing effect on microorganisms in the 
air, and at the same time has the odor-removal characteristics, 30minutes of 20mg/m³ (9.35PPM) 
concentration of ozone in a sealed environments can achieve disinfection effect. Ozone on the surface 
of the items of the microorganism also has a killing effect, requiring 60mg/m³ (28.04PPM), relative 
humidity of 70%, and exposure of 60 minutes to 120 minutes, to achieve disinfection effect.

2.  If the equipment is stopped in the middle of the operation, the disinfection effect cannot be guaranteed, 

a new disinfection session from the start will be necessary.

Summary of Contents for UVP401R

Page 1: ...V OZONE Sterilization Case PLUS UVC LED 120 275nm UVC Work light specifications Center box Irradiance 300μW cm Input MAX 25W Output 3 4h 4 2 kg IP65 60mg m 27ppm 10 95V 8000mAh Li ion UV 4 5h O3 15h 12 6V 4 3A 365x297x168mm ...

Page 2: ...amage of ultraviolet or ozone to human body With the intelligent button operation interface it is able to independently select and adjust the disinfection mode and disinfection time Color LCD screen can display the information e g disinfection mode disinfection time and state of charge etc in real time The external case is under the German Quality with strong corrosion resistance and shock resista...

Page 3: ...t long pressing it will automatically switch off after 30 seconds for energy conservation and environment protection Under the ultraviolet disinfection mode the white will change to purple Under the ozone disinfection mode the white will change to blue 6 If the disinfection case is open during the disinfection and the screen will display will disappear and the time will return to the preset time 7...

Page 4: ...t children before use and ensure that children do not touch the product 4 Avoid the risk of electric shock during use 5 The use of this product in gas flammable and explosive substances and heat source environments is prohibited 6 Don t use wet hands in operating the product Please keep the operating environment dry 7 Do not remove repair and alter the product by oneself Repairs and battery replac...

Page 5: ... pay full attention to the use of ozone 6 At Ozone concentrations of 0 02PPM 0 04mg m ozone can be smelled by people with acute olfactory senses at 0 15PPM 0 32mg m ozone can be smelled by the general population at 1 10PPM 2 14 21 4mg m ozone has reached the stimulus range and can be clearly smelled at 10PPM 21 4mg m ozone has reached the toxic range or more poisonous range The odor of ozone is no...

Page 6: ...ght source the distance of the disinfection object from the UV light source and the duration of the exposure to UV light 2 When killing common microorganisms the irradiated dose should reach 10 000 μW cm 100 000 μ W cm is the recommended dose to kill bacterial spores The radiation dose should not be less than 100 000 μW cm when the target microorganism for disinfection is unknown 3 According to th...

Page 7: ...leather material colors samples the results show color fading and hardening of the material especially the brighter samples whose fading was more pronounced 5 The same conditions were tested on stainless steel aluminum gold plated PIN needles silicone cotton pads cotton and ABS plastics and no significant decomposition was found 6 It is recommended to avoid ozone disinfection of brightly colored l...

Page 8: ...structions WARNING UV C emitted from this product Avoid eye and skin exposure to unshielded product Follow installation instructions and user manual www uvpluss com AEC Lighting Solutions Co Ltd Add 2548 Bao an Highway Jiading District Shanghai 201801 China Tel 400 083 8757 www aeclight reel com ...

Reviews: