DP 100 User’s Manual
22
6.2
Weight Calibration
Turn on and warm up the indicator for 15 to 30 minutes before calibration.
Adjust the calibration switch to “ON”. The screen will show
.
Press or to select
.
Procedure
Zero Calibration
a) Ensure there is nothing on the platform; after being stable, press , the screen will display
“
…….
”. The zero calibration will be completed 5 seconds later.
b) To abort zero calibration, just press .
Weight Calibration
a) Place an object with a known weight on the platform. Input the weight value from front panel.
After being stable, press . The screen will show “
…….
”. The weight calibration will be
completed 5 seconds later.
b) To abort weight calibration, just press .
Zero calibration completed
Confirmed
Set calibration
switch to “LOCK”
Place the
mass and
set its
weight
Zero
Calibration
Weight
Calibration
Empty Scale