background image

4.1

DISaSSEMBLy aND CLEaNING PROCEDURE

The  PIRVOL  should  be  inspected  periodically  to 
ensure that its performance is not impaired by fouled 
optics  or  by  clogging  of  the  sample  draw  cup  or 
hydrophobic  screen.    Inspection  and/or  periodic 
maintenance involves three different areas.

IMPORTANT

Remove  power  before  disconnecting  and 
removing the PIRVOL for maintenance.

Sample  Draw  Cup.   

Perform  a  visual  inspection  of 

the  sample  draw  cup  and  associated  plumbing,  and 
clean as required.  External filtration of sample stream 
is recommended.

hydrophobic  Screen.   

The  flow  of  gas  through  the 

screen  can  be  inhibited  by  an  accumulation  of  mud 
or  other  contaminants.    To  inspect  the  hydrophobic 
screen, disassemble the PIRVOL as described below.  
If the screen appears to be fouled, replace it.

Optics.   

Cleaning  of  the  optical  surfaces  is  required 

only  if  an  optical  fault  is  indicated  (1.0  milliampere 
current  output  signal  from  the  PIRVOL).    This 
procedure is most easily accomplished on a bench.

Required  materials:    Clean,  flat  work  surface, 
foam  tipped  swabs  (no  cotton),  isopropyl  alcohol, 
screwdriver.

CAUTION

The  PIRVOL  contains  semiconductor  devices 
that  are  susceptible  to  damage  by  electrostatic 
discharge.    An  electrostatic  charge  can  build 
up on the skin and discharge when an object is 
touched.  Therefore, use caution when handling 
the  device,  taking  care  not  to  touch  electronic 
components  or  terminals.    If  the  electronics 
assembly  is  removed,  it  should  be  placed  in  an 
anti-static bag or box while stored or transported.  
A  static  safeguarded  work  area  is  highly 
recommended (if available) for disassembly and 
cleaning of the PIRVOL.

1.  Remove the sample draw cup from the front of the 

PIRVOL.  See Figure 12.

2.  Perform  a  visual  inspection  of  the  sample  draw 

cup  and  associated  plumbing,  and  clean  as 
required.  Also inspect the two O-rings inside the 
sample draw cup and clean, lubricate or replace 
as required.

3.  Loosen  the  two  captive  screws  on  top  of  the 

mirror  assembly  (Figure  12)  and  remove  the 
mirror assembly and hydrophobic screen from the 
electronics assembly.

4.  Thoroughly  douse  the  interior  of  the  mirror 

assembly  as  well  as  the  foam  tipped  swab 
with  isopropyl  alcohol.    Use  the  swab  to  gently 
cleanse  the  surfaces  of  the  reflecting  mirrors 
inside  the  mirror  assembly.    After  cleaning  with 
the  swab,  flush  out  the  mirror  assembly  using 
a  liberal  amount  of  isopropyl  alcohol.    Tip  the 
mirror  assembly  with  mirror  openings  downward 
to  remove  accumulated  isopropyl  alcohol  and 
particle  contaminants.    Repeat  the  alcohol  flush 
to remove any remaining contaminants.  Allow the 
mirror assembly to air dry in a dust-free location.

IMPORTANT

Do  not  insert  any  sharp  object  into  the  mirror 
assembly.    Scratching  of  the  mirrors  will  void 
the PIRVOL warranty.  Do not use cotton tipped 
swabs  or  buds  as  they  are  likely  to  leave  fiber 
residue.

5.  Clean  the  two  windows  on  the  electronics 

assembly  using  the  procedure  described  above.  
When  the  mirror  assembly  and  electronics 
assembly  are  thoroughly  dry,  proceed  with  re-
assembly.

6.  Slide  a  new  hydrophobic  screen  over  the  end  of 

the mirror assembly, then carefully seat the mirror 
assembly  and  hydrophobic  screen  against  the 
base of the electronics assembly.  Be careful not 
to crumple or fold the hydrophobic screen.

7.  Tighten  the  two  captive  screws  on  the  top  of  the 

mirror assembly.  Tighten the screws evenly.  Do 
not over-tighten (6 inch-pounds recommended).

8.  Slide the sample draw cup back onto the PIRVOL.

9.  Check for leaks around the PIRVOL sample draw 

cup.

10

95-8522

Summary of Contents for DET-TRONICS 95-8522

Page 1: ...Instructions 95 8522 Volumetric IR Process Gas Monitor PIRVOL Series 4 1 7 07 95 8522 ...

Page 2: ...ay Device 5 Operating Modes 5 Installation 6 Device Mounting Orientation 6 Junction Boxes 6 General Wiring Requirements 6 Wiring Procedure 7 Detector Separation Optional 8 Startup Procedure 9 troubleshooting 9 Maintenance 9 Disassembly and Cleaning Procedure 10 Calibration 11 Device Repair and Return 12 Ordering Information 13 ...

Page 3: ...e CSA certified for use in Class I Division 1 Groups B C and D and Class I Division 2 Groups A B C and D areas They are fitted with a sealed sampling cup with compression type 0 25 inch o d tubing fittings for easy integration into a vapor sample extraction system FEATURES Routine calibration not required Continuous self test automatically indicates a fault or fouled optics condition Performs well...

Page 4: ...hane calibrated PIRVOL LT 95 8522 Figure 1 4 to 20 ma Current Loop Resistance Table 1 Current Loop Output Levels and Corresponding Status Indications Current Level Status 23 0 ma 100 by volume 20 0 ma 90 by volume 4 0 ma Zero gas level 0 by volume 2 2 ma Zero calibration in progress 2 0 ma Span calibration in progress 1 8 ma Calibration complete remove gas 1 6 ma Calibration fault 1 0 ma Fouled op...

Page 5: ...h 5 watt walkie talkie keyed at 1 meter ENCLOSURE MATERIALS Copper free aluminum clear anodized enclosure Stainless Steel 316 electropolished enclosure SAMPLE DRAW CUP Material Aluminum Fittings Brass Tubing Size 1 4 inch O D MOUNTING Det Tronics junction box available in a tall cover with window or a short cover version is recommended for optimum ease of installation PIRVOL can be threaded into a...

Page 6: ...s designated reference and the other active with different optical filters in front of the two sensors to make them selective to different infrared wavelengths The reference wavelength is unaffected by combustible gases while the active wavelength is absorbed by combustible gases The ratio of the active to the reference wavelength is computed within the PIRVOL to determine the concentration of gas...

Page 7: ...PIRVOL it enters a Warmup mode for approximately one minute in which it performs diagnostic checks and allows the sensors to stabilize before beginning normal operation The current output during this period is 0 milliamperes At the end of the warmup period with no faults present the PIRVOL automatically enters the Normal operating mode If a fault is present after the warmup the current output will...

Page 8: ...th the PIRVOL Tall Cover Model junction box cover includes a window See Figure 6 Short Cover Model junction box uses solid cover This junction box can also be used for detector separation See Figure 7 General Wiring Requirements NOTE The wiring procedures in this manual are intended to ensure proper functioning of the device under normal conditions However because of the many variations in wiring ...

Page 9: ... 4 to 20 ma signal output Yellow lead Calibration input Green lead Chassis ground If the calibration wire yellow lead is not being used do not connect this wire to ground Trim excess length and insulate wire so no shorting can occur 95 8522 Figure 6 Tall PIRVOL Junction Box with Window Figure 7 Short PIRVOL Junction Box Figure 9 Typical PIRVOL Wiring with Det Tronics Junction Box DET TRONICS JUNCT...

Page 10: ... the transmitter and then to the PIRVOL to ensure that the power requirements are met Mounting and Connecting Procedure for Detector Separation The PIRVOL junction box can be mounted to a wall or post or it can be suspended by the conduit if this does not result in excessive vibration The junction box should be electrically connected to earth ground 1 Lubricate the sensor threads with low vapor pr...

Page 11: ...ng threads on the PIRVOL and associated junction box A suitable grease is listed under Spare Parts in the Ordering Information section of this manual It is recommended to keep spare IR modules See Spare Parts section for field replacement in the event of a malfunction 95 8522 Current Level Status Corrective Action 2 4 to 3 9 ma Negative Zero Indication NOTE This fault can be caused by the presence...

Page 12: ...up from the front of the PIRVOL See Figure 12 2 Perform a visual inspection of the sample draw cup and associated plumbing and clean as required Also inspect the two O rings inside the sample draw cup and clean lubricate or replace as required 3 Loosen the two captive screws on top of the mirror assembly Figure 12 and remove the mirror assembly and hydrophobic screen from the electronics assembly ...

Page 13: ...rocured Calibration Procedure To perform a zero calibration of the PIRVOL follow the sequence described in Table 3 NOTE The calibration sequence is initiated by momentarily connecting the calibration lead to the negative lead common of the power supply using either the Cal Magnet or an external switch If the Det Tronics junction box with magnetic Cal Switch is being used this is accomplished by ho...

Page 14: ... packing material in addition to an anti static bag or aluminum backed cardboard as protection from electrostatic discharge Return all equipment transportation prepaid to the factory in Minneapolis Description Current LED Operator Action Normal operation no gas present 4 0 ma Off If the possibility of background gases exists purge the sensor with clean air to ensure accurate calibration Initiate c...

Page 15: ...7128 001 Sample Draw Cup with O Rings and Fittings 007128 002 O Ring Kit UL 006297 003 Hydrophobic Screen 006680 001 Silicone Grease Assistance For assistance in ordering a system to meet the needs of a specific application contact Detector Electronics Corporation 6901 West 110th Street Minneapolis Minnesota 55438 USA Operator 952 941 5665 or 800 765 FIRE Customer Service 952 946 6491 Fax 952 829 ...

Page 16: ... IR Combustible Gas Detector Eagle Quantum Premier Safety System Eagle Logic Solver Safety System Det Tronics the DET TRONICS logo Eagle Quantum Premier and Eclipse are registered trademarks or trademarks of Detector Electronics Corporation in the United States other countries or both Other company product or service names may be trademarks or service marks of others Copyright Detector Electronics...

Reviews: