insufficient at the end of the fill/evac stage, the test is aborted as a gross failure.
Fine leaks are determined with the force decay in the measure step.
Measure type:
Defaults to force change.
Test Pressure:
The
pressure the test is set at. Used as a guide for defaults and can
be changed later in the programming.
5.
Configure the Steps for the new program
by first selecting the
Specifications
tab,
then selecting/configuring the desired
Program number
,
Test type
,
Test pressure
,
and
Test tolerances
.
6.
Press
Apply Defaults
. The QMR will auto-populate the required steps and default
times into the steps screen. See Figure 14.
7.
After “Applying Defaults” to the selected program, select the
Step
tab (Figure 15) to
finish configuring step parameters, as described below:
Figure 15: Steps Tab
Figure 14: Test Setup/Specifications Tab
Apply Defaults
Note: Test Volume should be the size
of the chamber to ensure correct fill
time.