18
DC Current
Range
Resolution
Accuracy
400µA
0.1µA
± (1.2% + 3)
4000µA
1µA
40mA
0.01mA
± (1.5% + 3)
400mA
0.1mA
4A
0.001A
± (1.8% + 3)
10A
0.01A
± (2.0% + 5)
Overload Protection:
Fuse 1 : F 400mA/690V - Fuse 2: F 10A/690V
Max. Input Current:
10A ( For inputs > 2A : measurement duration
< 10 secs, interval > 15 minutes )
Max. Measurement Voltage Drop:
400mV
Note:
The 10A range is specified from 20% to 100% of range.
AC Current
Range
Resolution
Accuracy
400µA
0.1µA
± (1.5% + 5)
4000µA
1µA
40mA
0.01mA
± (1.8% + 5)
400mA
0.1mA
4A
0.001A
± (2.0% + 5)
10A
0.01A
± (2.5% + 10)
DC Voltage
Range
Resolution
Accuracy
400mV
0.1mV
± (1.0% + 5)
4V
0.001V
± (0.8% + 3)
40V
0.01V
400V
0.1V
600V
1V
± (1% + 5)
Input Impedance:
the range 400mV: >1000M
ȍ
/ the other ranges: 10M
ȍ
Overload Protection:
600V DC/AC rms
AC Voltage
Range
Resolution
Accuracy
4V
0.001V
± (1.0% + 5)
40V
0.01V
400V
0.1V
600V
1V
± (1.2% + 5)
Input Impedance:
10M
ȍ
Frequency Range:
40Hz -400Hz
Response:
Average, calibrated in rms of sine wave
Overload Protection:
600V DC/AC rms