![US Ultratek USB-UT350 User Manual Download Page 9](http://html1.mh-extra.com/html/us-ultratek/usb-ut350/usb-ut350_user-manual_1063781009.webp)
USB-UT350 User’s Guide
Revision 1.77 – 09/30/2009
4-2
4.5 DC
Offset
Sets the DC offset of the signal.
4.6 Buffer
Length
The buffer length specifies how many samples the USB-UT350 will analyze during data
acquisition. This value can be anywhere between 2 and 8191 samples,
or 1 to 32000 with
32K memory option
.
4.7 Trigger
Delay
The trigger delay specifies how many samples the A/D converter should skip before
processing the response signal. This value can be anywhere between 0 and 16,383
samples.
4.8 Average
Sets the signal averaging. The device will take multiple waveforms and then average
them.
4.9 Polarity (Tone Burst Pulser Only)
Sets the polarity of the tone burst pulser. Positive polarity starts the oscillation with a
positive pulse while negative polarity starts the oscillation with a negative pulse.
4.10 Signal Type
There are four types of signal available:
Signal Type
Application
Full
Full wave rectification
Flaw detection
+Half
Positive half rectification
Thickness measurement
-Half
Negative half rectification
Thickness measurement
RF
No rectification
Thickness measurement