background image

 

Z650HR STEREO MICROSCOPE  

 

 

 
 

 

UNITRON

®     

                          

73 Mall Drive, Commack, NY  11725  •  631-543-2000  •  www.unitronusa.com 

8

 

ADJUSTMENT  
AND OPERATION 

Adjusting Trinocular Viewing Head 
 

 

 

Adjusting Interpupillary Distance (Fig. 6)  

Different users have different interpupillary distances 
(this distance is between the centers of the pupils of 
each eye). When the operator changes, it will be 
necessary to adjust the interpupillary distance. 

While looking through the eyepieces, hold the left 
and right eyetubes of the viewing head

 

and adjust the 

eyetubes by opening or closing them until the left and 
right fields of view coincide completely and you are 
able to see a complete circle. 

Adjusting Diopter Ring (Fig. 7) 

 

Set the dioptor rings

 ② 

of both eyepieces to “0” 

position. (Do this when users change, because 
different users will have different diopter settings.) 

Place an easy-to-observe specimen on the stage 
plate, i.e., a coin. 

Rotate the zoom knob

 ③ 

to the highest 

magnification 5x, then turn the focusing knob

 ④ 

to 

focus the specimen.

  

Rotate the zoom knob

 ③ 

to the lowest magnification 

0.8x, looking only into the left eyepiece, adjust the 
diopter ring on left eyepiece to focus the specimen. 
Then repeat procedure for the right eyepiece.  

NOTE:

  The working distance (the distance between 

the microscope objective to the top of the specimen) 
of the microscope is 115mm, the trinocular head 
bracket/focus mount

 ⑤ 

can be mounted onto higher 

or lower positions according to your needs by 
removing the clamping bolt 

 (Fig. 9 next page) 

with the Allen wrench provided. 
 

Third Optical Path Parfocality Adjustment (Fig. 8) 

 

After adjusting the diopter, the image should be 
clear both in the eyepiece and the camera.  If the 
image is not clear in the camera, remove the rubber 
screw cover 

, loosen the inner screw and rotate 

the screw with a flat head screwdriver until the 
image is clear.  When clear, tighten the screw and 
replace the screw cover.

 

Fig. 6 

Fig. 7 

Fig. 8 

Summary of Contents for Z650HR Series

Page 1: ...73 Mall Drive Commack NY 11725 631 543 2000 P 631 589 6975 F www unitronusa com info unitronusa com Z650HR SERIES ZOOM STEREO MICROSCOPE MANUAL ...

Page 2: ...NTENTS SAFETY NOTES 3 CARE AND MAINTENANCE 3 INTRODUCTION 4 UNPACKING AND COMPONENTS 4 COMPONENTS DIAGRAM 4 ASSEMBLY 5 DETAILED ASSEMBLY PROCEDURE 6 ADJUSMENT AND OPERATION ADJUSTING BINOCULAR VIEWING HEAD 8 FOCUS ADJUSTMENT 9 SPECIFICATIONS 10 TROUBLESHOOTING 11 MAINTENANCE 12 SERVICE 12 WARRANTY 12 ...

Page 3: ...e NOTE Always plug the microscope power cord into a suitable grounded electrical outlet A grounded 3 wire cord is provided CARE AND MAINTENANCE 1 Do not attempt to disassemble any component including eyepieces objectives or the focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persistent dirt s...

Page 4: ...l procedures ensure each microscope is of the highest quality prior to shipment UNPACKING AND COMPONENTS Your microscope arrived packed in a molded shipping carton Do not discard the carton the shipping carton should be retained for reshipment of your microscope if needed Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew can form Carefully ...

Page 5: ...EMBLY The diagram below shows how to assemble the various modules The numbers indicate the order of assembly When assembling the microscope make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces Eyepieces Trinocular Head Power Cord Stand Lock screw ...

Page 6: ...2 Fig 3 Remove the eyepiece dust caps Fig 1 and gently insert the two adjustable eyepieces Fig 2 into the eyepiece sleeves until they stop as shown in Fig 3 Replace the dust caps when not in use IMPORTANT ENVIRONMENT REQUIREMENTS 1 Ambient temperature 0 C to 40 C 32 F to 104 F Maximum R H 85 2 High temperature will result in moisture and mildew and could damage instrument 3 Avoid placing the instr...

Page 7: ... of a different input voltage than indicated will cause severe damage to your microscope Connecting the Power Cord Fig 4 Turn the power switch to the off position O before connecting the power cord Insert the power cord plug into the power jack of the microscope make sure the connection is snug Plug the power cord into the power supply receptacle Using the Glass Stage Fig 5 Set the glass stage pla...

Page 8: ...ve specimen on the stage plate i e a coin Rotate the zoom knob to the highest magnification 5x then turn the focusing knob to focus the specimen Rotate the zoom knob to the lowest magnification 0 8x looking only into the left eyepiece adjust the diopter ring on left eyepiece to focus the specimen Then repeat procedure for the right eyepiece NOTE The working distance the distance between the micros...

Page 9: ...djustment has been completed always rotate both adjustment knobs in the same direction Changing the Magnification The zooming knobs located on both sides of the zooming body will change the magnification of the specimen image Total Magnification Magnification of zoom body x magnification of eyepiece i e 0 6 x 10 6x NOTE If using any of the optional auxiliary objectives then the above magnification...

Page 10: ... LED lamp illumination Input 100 240V 0 15A 50 60Hz Auxiliary Objective Working Distance mm Eyepiece WF10X 23 WF15X 16 WF20X 12 WF30X 18 Total Magnification Visual Field mm Total Magnification Visual Field mm Total Magnification Visual Field mm Total Magnification Visual Field mm 115 6 50X 38 3 4 6 9 75X 26 7 3 2 12 100X 20 2 4 18 150X 13 3 1 6 0 5X 220 3 25X 76 7 9 2 4 5 37 5X 53 3 6 4 6 50X 40 4...

Page 11: ...ges Interpupillary distance is not correct Readjust it Diopter adjustment is not correct Readjust it Dirt appears in the view field Dirt on the specimen Clean specimen Dirt on the surfaces of eyepieces Clean eyepieces Image is not clear Dirt on the surface of objectives Clear objectives Image is not clear while focusing change Diopter adjustment is not correct Readjust diopter Focusing is not corr...

Page 12: ...om defects in material and workmanship for a period of five 5 years for mechanical and optical components and one 1 year for electrical components from the date of invoice to the original end user purchaser This warranty does not cover damage caused in transit misuse neglect abuse or damage resulting from improper servicing or modification by other then UNITRON approved service personnel This warr...

Reviews: