Unitron RMM2 Manual Download Page 9

 

RMM2 ROLLSCOPE MICROSCOPE 

 

 

 

 
 

UNITRON

®     

 

                          

73 Mall Drive, Commack, NY  11725  •  631-543-2000  •  www.unitronusa.com 

 

 
 
Using the Field Diaphragm 

 

While observing a specimen that has been focused, slowly open and close the field diaphragm using the 
lever on the illuminator assembly.  The diaphragm is used to eliminate glare by restricting the specimen 
illumination to the area actually being observed.  Therefore, the diaphragm should be opened only until 
the full field of view is seen completely illuminated and no more.  Opening the diaphragm more may 
cause glare that will produce “washed-out” images. 
 
 
 

Visual Observation 

 

The top eye lens of the WF 10x eyepiece has a threaded focusable mount.  Before focusing the 
specimen, hold the lower part of the eyepiece with one hand and rotate the upper threaded focusable 
eye lens with the other hand until the image of the eyepiece reticle is in focus.  Rotate the coarse focus 
knob to obtain approximate focus of the specimen using the 5x objective.  Then rotate the fine focus 
knob until the specimen is in sharp focus. A slight adjustment with the fine focus knob may be necessary 
when changing objective power. 
 
 
 

Changing Objective Magnifications 

 

Total magnification is the product of the eyepiece power multiplied by the power of the objective.  The 
objectives are parfocal.  Therefore, once an image is in sharp focus, only a slight adjustment of the fine 
focus control will be needed when rotating the nosepiece from one objective power to another. 
 
 
 

Filters 

 

Two filters, green and yellow, are supplied in metal mounts with handles.  These fit into the illuminator 
filter slot.  The filters are used to create contrast in the specimen, minimize reflections and produce 
images that are most satisfactory to the observer.  The achromatic objectives are designed to give the 
highest degree of color correction and maximum image quality when used with green light, the color to 
which the human eye is most sensitive.  Therefore, the green filter is recommended for most visual 
observations. 
 
 
 

Measuring Reticle 

 

The crossline reticle is already installed into the eyepiece.  The reticle, when magnified and seen through 
the microscope, is approximately equal to the ruled value of the reticle divided by the magnification of the 
objective. 
 
 

 

Summary of Contents for RMM2

Page 1: ...73 Mall Drive Commack NY 11725 631 543 2000 P 631 589 6975 F www unitronusa com info unitronusa com RMM2 ROLLSCOPE MICROSCOPE MANUAL ...

Page 2: ... 631 543 2000 www unitronusa com 2 CONTENTS SAFETY NOTES 3 CARE AND MAINTENANCE 3 INTRODUCTION 4 UNPACKING AND COMPONENTS 4 COMPONENTS DIAGRAM 6 ASSEMBLY 7 OPERATION 7 PHOTOMICROGRAPHY 10 ADJUSTMENT AND CARE 11 SPECIFICATIONS 12 MAINTENANCE 13 SERVICE 13 WARRANTY 13 ...

Page 3: ...OTE Always plug the stereomicroscope power cord into a suitable grounded electrical outlet A grounded 3 wire cord is provided CARE AND MAINTENANCE 1 Do not attempt to disassemble any component including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persistent dirt sh...

Page 4: ...power objectives as well as in choosing and composing areas for photography UNPACKING AND COMPONENTS Your microscope arrived packed in a molded shipping carton Do not discard the carton the shipping carton should be retained for reshipment of your microscope if needed Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew can form Carefully remo...

Page 5: ...t photographs are being made PLANE GLASS REFLECTOR No adjustment is necessary to provide ideal conditions for vertical illumination The plane glass reflector has been factory installed adjusted and inspected and is correctly positioned to assure even full field illumination No positioning change is required before use unless the reflector has been disturbed in transit assembly or usage DIAL DEPTH ...

Page 6: ...icrometrics Digital Camera Depth Gauge Depth Gauge Bracket With Lock Screw X Axis Control Graduated Positioning Scale Fine Focus Knob Roll Feet Contact Screw Vertical Illuminator Not shown Bulb cord and socket assembly Bulb Centering Screws Y Axis Control Coarse Focus Knob Field Diaphragm Control Lever Illuminator Filter Slot Objectives Nosepiece Optional Accessory ...

Page 7: ...life In general use higher illumination intensities to provide short exposure times during photography or for examining poorly reflecting materials Always turn the transformer off when the microscope is not in use Operation With Batteries The illuminator may be operated by batteries for completely portable operation NOTE Insert the metal battery spacer into each foot before inserting the batteries...

Page 8: ... tightened or loosened The bulb position is optimum when the filament image appears as shown in Figure C below Figures A and B illustrate an off centered filament Lamp Filament Alignment 2 Referring to the figure above note that there are three conditions to be met a The full square area of the filament should be seen This is obtained by rotating the bulb socket in the sleeve of the centering adju...

Page 9: ...imen is in sharp focus A slight adjustment with the fine focus knob may be necessary when changing objective power Changing Objective Magnifications Total magnification is the product of the eyepiece power multiplied by the power of the objective The objectives are parfocal Therefore once an image is in sharp focus only a slight adjustment of the fine focus control will be needed when rotating the...

Page 10: ...auge reading gives a direct measurement of the difference in height between the two levels The highest measuring accuracy will be obtained by using the 50x objective since this objective has the smallest depth focus NOTE This procedure is not necessary with the digital depth gauge X Y Motion The RMM2 Series is equipped with a built in mechanical bench which slides the microscope body in X Y direct...

Page 11: ...t cover The eyepiece should always be in the microscope to eliminate the possibility of dirt or dust from entering the vertical tube and falling onto the back lenses of the objectives Should the optics require cleaning only use the finest grade of soft camera lens tissue If foreign matter or stubborn dirt has adhered to a lens surface a small amount of xylene applied sparingly can be applied with ...

Page 12: ...tor Variable 6 volt 20 watt Koehler type vertical halogen illuminator with field diaphragm for both AC and DC operation Filters Green and yellow filters are standard A filter slot is built into the illuminator housing Dimensions Weight Physical weight dimensions 9 lbs 9 x 11 x 10 WDH Shipping weight and dimensions 28 lbs 21 x 14 x 16 Warranty 5 year limited warranty instruction manual and dust cov...

Page 13: ...defects in material and workmanship for a period of five 5 years for mechanical and optical components and one 1 year for electrical components from the date of invoice to the original end user purchaser This warranty does not cover damage caused in transit misuse neglect abuse or damage resulting from improper servicing or modification by other then UNITRON approved service personnel This warrant...

Reviews: