MEC2 INVERTED METALLURGICAL MICROSCOPE
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com
13
Adjusting the Diopter
(Figure 21)
1. Looking through the right ocular only with your
right eye, use the coarse and fine focusing
adjustment knobs to focus on the specimen.
2. Then with your left eye only, look through the
left ocular. If the image is not sharp, rotate the
diopter adjustment ring
①
to focus on the
specimen.
Diopter adjustment range: ±5
Adjusting the Interpupillary Distance
(Figure 22)
While observing with both eyes, hold the left and
right prism holders (2 & 3), rotate them around
the axis to adjust the interpupillary distance until
the left and right fields of view coincide to form a
single image
.
Interpupillary range
:
48
~
75mm
Switching the Light Path
(Figure 23)
Pull out or push in the light path selector lever
(4) to select the light path required.
For the binocular observation, push the lever in
completely .For video or microphotography, pull
the lever out completely.
Fig. 21
Fig. 22
Fig. 23