United Equipment EzFlow Installation & Maintenance Manual Download Page 9

  Page 

9

UNITED EQUIPMENT 

SALES CONSULTING CORPORATION

INSTALLATION OF THE MANIFOLD

1.  Remove contents from box and confirm that all required equipment has been supplied. Do not discard 

packing material until the system is in full operation. Keep all manuals and warranty cards.

2.  Remove manifold from box and inspect for damage.
3.  Open cabinet door to access manifold interior.
4.  Make sure that there are not any wires that are disconnected from their terminals.
5.  Using the manifold, mark mounting screw locations on the wall and install wall anchors if required  

 

 

NOTE:

 Use appropriate wall anchors for a secure installation.

6.  It is important to centrally locate the manifold amongst all four cylinders to ensure that all hoses can 

reach the appropriate connections.

 

NOTE:

 Position the manifold at a height that prevents any interference with the cylinders.

7.  Using both the Oxygen and Nitrous Oxide pipes, connect the manifold line ports (at the top of the 

manifold) to the facility's main gas lines. Pipeline connections have been supplied to be brazed into the 

main facility line (follow all local and national codes applicable to your area).

8.  Install the tank restraint brackets on the wall as per local codes. Place cylinders in tank restraints and 

secure. Make certain all valves have been closed.

 NOTE:

 Before connecting regulators to cylinder, momentarily open and close cylinder valve to blow out  

  dirt and debris.

9.  Install gas specific regulators in cylinders intended for operation. Be certain not to over tighten. Please 

check with your medical gas supplier to determine the correct torque needed to tighten the connection.

10. Using the two Oxygen and two Nitrous Oxide 6' (1.8 m) hoses, connect the appropriate hoses to the 

input lines located on the bottom of the manifold gas blocks. It is recommended to align cylinders on the 

wall in relation to the inputs on the manifold. This will allow for easier identification of tanks during setup/

change-out.

11. Connect the loose end of the hoses to the gas specific DISS connection on the regulators.
12. Locate the RAC and connect using a CAT5 or CAT6 cable (cable supplied by others). See page 10.
13. Open all cylinders S-L-O-W-L-Y and check for leaks using Snoop Leak Detector.  

 

 

NOTE:

 Electrical power is required to operate the solenoids and restrict flow. Without electrical power,   

  pressure will fill the lines.

14. Plug in manifold to a 120 V-240 V power supply (do not use extension cords).
15. Confirm that the manifold is on. 

INSTALLATION 

INSTRUCTIONS

Summary of Contents for EzFlow

Page 1: ...INSTALLATION MAINTENANCE MANUAL AUTOMATIC MANIFOLD...

Page 2: ...troller 8 Installation Instructions 9 10 Installation of the Manifold 9 Installation of the Remote Alarm Controller 10 Regulator Settings 11 Display Functions 11 Testing 12 15 Testing for N2O Emergenc...

Page 3: ...damage to equipment DANGER denotes steps which can prevent electrical shock to equipment or to prevent serious injury and or death M i c r o p r o c e s s o r D i g i t a l M a n i f o l d USER RESPON...

Page 4: ...ch to the secondary reserve cylinder without interrupting the supply of gas to the facility The red LED will illuminate and an alarm will sound when a bank has depleted and requires replacement The Ez...

Page 5: ...onformity with the descriptions contained in this manual when assembled operated maintained and serviced in accordance with the installation instructions provided The manifold must be checked periodic...

Page 6: ...n the primary bank of cylinders is depleted the manifold will automatically switch to the secondary bank of cylinders without interruption of gas flow to the facility The red LED will illuminate when...

Page 7: ...ve pressure in both the manifold and the line by using the bleed valve prior to removing or servicing the sensor cover circuit board There are two line pressure sensors installed within the manifold F...

Page 8: ...henever the system is not in use This ensures that a dangerous condition will not occur if the lines are left open Simply press the Stand By button to return the manifold to full operation NOTE If equ...

Page 9: ...rackets on the wall as per local codes Place cylinders in tank restraints and secure Make certain all valves have been closed NOTE Before connecting regulators to cylinder momentarily open and close c...

Page 10: ...C from packaging 4 Loose set screw do not discard located at bottom of box and remove front cover Set cover aside do not discard 5 Plug the cable running from the manifold circuit board into the back...

Page 11: ...not relieved prior to disconnection M i c r o p r o c e s s o r D i g i t a l M a n i f o l d P a g e 4 USER RESPONSIBILITY The information contained in this Installation and Maintenance Manual perta...

Page 12: ...the LED display will read LO 6 At this time the nitrous oxide LED display will show OFF and will flash showing that the nitrous oxide has been closed due to lack of oxygen in the system An alarm will...

Page 13: ...wing the cylinder pressure to drop Once the pressure gets below the low set point the system will change over to the reserve tank previously represented by the yellow LED The closed cylinder will show...

Page 14: ...STAND BY 1 To place the system in stand by mode push and hold the Stand By button for five seconds on either the manifold or the RAC This safety feature prevents the system from entering the stand by...

Page 15: ...operation until a qualified person has inspected and passed the installation as per NFPA 99 or other local standards M i c r o p r o c e s s o r D i g i t a l M a n i f o l d P a g e 4 USER RESPONSIBI...

Page 16: ...tributors Installing CO2 and N2O manifolds outdoors Please refer to NFPA Code 5 1 3 3 1 8 Central supply systems for nitrous oxide and carbon dioxide shall be prevented from reaching temperatures lowe...

Page 17: ...e used shall be Belden 8451 or equivalent 22 gauge shielded twisted pair for multiple sensors a multi conductor shielded twisted pair cable should be used CONNECTING THE REMOTE SENSOR WIRES TO THE MAN...

Page 18: ...Page 18 UNITED EQUIPMENT SALES CONSULTING CORPORATION NOTES...