INSTALLATION OF THE MANIFOLD
1. Remove contents from box and confirm that all required equipment has been supplied. Do not discard
packing material until the system is in full operation. Keep all manuals and warranty cards.
2. Remove manifold from box and inspect for damage.
3. Open cabinet door to access manifold interior.
4. Make sure that there are not any wires that are disconnected from their terminals.
5. Using the manifold, mark mounting screw locations on the wall and install wall anchors if required
NOTE:
Use appropriate wall anchors for a secure installation.
6. It is important to centrally locate the manifold amongst all four cylinders to ensure that all hoses can
reach the appropriate connections.
NOTE:
Position the manifold at a height that prevents any interference with the cylinders.
7. Using both the Oxygen and Nitrous Oxide pipes, connect the manifold line ports (at the top of the
manifold) to the facility's main gas lines. Pipeline connections have been supplied to be brazed into the
main facility line (follow all local and national codes applicable to your area).
8. Install the tank restraint brackets on the wall as per local codes. Place cylinders in tank restraints and
secure. Make certain all valves have been closed.
NOTE:
Before connecting regulators to cylinder, momentarily open and close cylinder valve to blow out
dirt and debris.
9. Install gas specific regulators in cylinders intended for operation. Be certain not to over tighten. Please
check with your medical gas supplier to determine the correct torque needed to tighten the connection.
10. Using the two Oxygen and two Nitrous Oxide 6' (1.8 m) hoses, connect the appropriate hoses to the
input lines located on the bottom of the manifold gas blocks. It is recommended to align cylinders on the
wall in relation to the inputs on the manifold. This will allow for easier identification of tanks during setup/
change-out.
11. Connect the loose end of the hoses to the gas specific DISS connection on the regulators.
12. Locate the RAC and connect using a CAT5 or CAT6 cable (cable supplied by others). See page 10.
13. Open all cylinders S-L-O-W-L-Y and check for leaks using Snoop Leak Detector.
NOTE:
Electrical power is required to operate the solenoids and restrict flow. Without electrical power,
pressure will fill the lines.
14. Plug in manifold to a 120 V-240 V power supply (do not use extension cords).
15. Confirm that the manifold is on.