CO-DV-IN-001 – Version 01 – 14 AUG 2020 

 

-

 

If the 

touchscreen

 is 

not responsive

, connect a mouse using a USB port and: 

o

 

Proceed  with  the 

calibration  of  the  touchscreen

  (follow  the  instructions 

provided in the User Manual). 

o

 

If the problem persists after the calibration of the touchscreen, proceed with 

the 

recovery of the touchscreen

 (follow the instructions provided in the User 

Manual). This option will restore the default settings of the touchscreen. 

 

-

 

If  the 

error  messages  E902

  (strong  light  intensity)  or 

E903

  (weak  light  intensity) 

appear, please  proceed with the  cleaning of  the  drawer  and the  white patch.  If  the 

messages appear after the cleaning, proceed with the calibration of the device using 

APP105 as explained previously. 

 

-

 

Upon any other error or improper performance of the device, which cannot be solved 

following all the mentioned procedures, it is recommended to conduct a Factory Reset 

of  the  instrument.  This  option  will  restore  the  default  settings  of  the  device. 

WARNING

: all the test data and method files saved in the instrument will be cleared, 

please export your data before!

 

 

Once the previous steps have been verified, if your problem is still not solved, please submit 

your complaint through 

[email protected]

 

Please be sure to transfer us the following information, in order to speed up the process of 

analysis of your submission: 

-

 

Serial Number

 of the instrument. 

-

 

Software version

 (both the 

embedded software

 and the 

PC manager software

). 

-

 

Result of the 

last calibration check

 (please provide a picture). 

-

 

Any  other  picture  which  might  be  relevant  for  the  diagnosis  and  the  root  cause 

analysis. 