Grounding may help reducing electrical noise when working in environments with high levels of such 
noise.  
 
Note: Never use the grounding cable with pH and redox sensors. 
 
For grounding the measuring setup, the banana plug is plugged into the connector on the back of 
the instrument and the silver wire is dipped into the liquid where the sensor is inserted (calibration 
standard, sample, water bath etc.). This will remove any difference in electrical potential between 
the liquid and the sensor.  
 
The PC must be connected to a proper grounding through its power supply and the mains cable. 
Make sure that the mains supply is properly grounded. 
 
NOTE: Some microsensors may only be used if the grounding cable is used. These sensors are 
marked 

Grounding cable must be used to keep warranty!

  In this case the grounding must be 

established as described above before the sensor touches the liquid and the cable must remain in 
place as long as the sensor is used in the liquid.   
 

8.3

 

Software 

The UniAmp instruments require the latest version of the SensorTrace Suite software. Check the 
version of your software in any of the applications (Logger, Profile, Rate, Programming Tool, Photo) 
by clicking 

Help

 - 

About

, or right click the UniAmp service in the 

Notification Area

 and select 

About.

 If 

you are not already running the latest version of the SensorTrace Suite, please download 
(

https://www.unisense.com/Software_download/

and install it (see 

6.1 Installation of the 

SensorTrace Suite software

).  

9

 

Analog in and out 

The Analog in/output is located on the back of the instrument (Figure 5). Using an adaptor, it is 
possible to retrieve the analog signals from all sensors for use with external analog to digital 
converters. The Analog out must be activated in the UniAmp service window (Figure 3). Contact 

[email protected]

 if you need the adaptor. For connection to non-Unisense equipment, contact 

Unisense first, to avoid damage to the instrument.    
 

 

Figure 5: Connectors for Analog and Digital in and output and for the grounding cable (right).  

10

 

Digital in and out 

For future implementation and for service purposes.  

Summary of Contents for UniAmp fx-3

Page 1: ...Unisense UniAmp Multichannel Systems User Manual UNIAMP MULTICHANNEL SYSTEM MANUAL Copyright Unisense A S Version 1 1 August 2020...

Page 2: ...ting the sensor polarization 6 7 4 Adjusting the offset 7 7 5 Setting STOX sensor parameters 7 7 6 Polarizing amperometric sensors without a PC connected 8 8 Sensor connections and compatibility 8 8 1...

Page 3: ...ation For further information please see the document General terms of sale and delivery of Unisense A S as well as the manuals for the respective products 3 Support ordering and contact information I...

Page 4: ...fication Area also called System tray in the lower right hand area of the screen Using this pop up you can adjust channel settings such as pre amplification offset sensor polarization etc 5 Operating...

Page 5: ...e drivers needed for connection will automatically be installed The Unisense UniAmp service is installed together with the SensorTrace Suite Click on the UniAmp service icon to open a window with a fr...

Page 6: ...sor signal in pico ampere pA is converted into a signal in millivolt mV This is set by the E2 PROM in the sensor when the sensor is plugged in but can be changed manually Simply select the Pre Amp Ran...

Page 7: ...revent O2 from reaching the measuring O2 sensor The measuring signal with the front guard on is thus the signal for zero oxygen The measuring signal with the front guard off is thus the signal for O2...

Page 8: ...or Manual To be downloaded at https www unisense com manuals 8 Sensor connections and compatibility 8 1 Sensors connectors and adaptors Sensors for the UniAmp meters must have the 10 pin UniAmp connec...

Page 9: ...is used in the liquid 8 3 Software The UniAmp instruments require the latest version of the SensorTrace Suite software Check the version of your software in any of the applications Logger Profile Rat...

Page 10: ...ure that you enter manually The input from a temperature sensor This sensor must be selected from the dropdown menu on the calibration window for each amperometric sensor The temperature correction of...

Page 11: ...Update Microsoft Windows Update may interfere with logging for longer periods of time Windows may force installation of updates and restart the PC This will obviously stop any logging of data in Sens...

Page 12: ...et and that it is not running on battery In the power saving settings on you PC disable all power saving options Please note that it may be necessary to disable power saving options both in Windows se...

Page 13: ...this type will often be periodic e g noise emitted when equipment is switching on or off If such problems occur try switching nearby equipment off one by one to identify which equipment is generating...

Reviews: