Function Description Related to Records
47
Chapter
7
7-2.
Data recording method
One-push
The indicated value is recorded to the point data at the timing the
key was pressed.
One-shot input
The indicated value is recorded to the point data at the timing when a short-circuit is detected in the
2-pin (TRG1) and COM in the external I/O connector.
Command input
The indicated value is recorded to the point data at the timing the RS-232C interface is used and a
CR command is received.
Level input
The indicated value continues to be recorded to the point data at the timing of indicated value
update (10 times per second) while a short-circuit is detected in the 3-pin (TRG2) and COM in the
external I/O connector.
Alarm detection
Indicated value in between seven seconds before (70 points) and three seconds after (30 points) in
alarm detection is recorded as alarm detection data.
7-3.
Alarm detection data, point data clear
Data and point data recorded upon alarm detection can be deleted from the main unit.
The alarm status is cleared by deleting the alarm detection data recorded upon alarm detection.
<<Setting value>>
ESC:
Cancel
CLr-AL:
Delete alarm detection data
CLr-Pt:
Delete point data
How to clear alarm data and point data
1.
Press the
key.
2.
Select the data to delete.
Press the
keys and select the
data to delete (initial value: ESC).
Press the
key to confirm and return
to the indicated value display.
* Press the
key to cancel the setting and return to indicated value display.
* When the alarm detection is set, the alarm will not be detected for approximately seven seconds
after returning to the indicated value display.
Attempt it after the REC status lights on.
REC
USB
ALARM
RST/.
V
A
VA
var
W
rpm
PF
Nm
V
A
VA
var
W
rpm
PF
Nm
V
A
VA
var
W
%
PF
V
A
°C
rpm
V
°C
Nm
1
3
2
Σ
1
3
2
Σ
1
3
2
Σ
1
AUX
2
Σ
ALARM
REC
FNC
USB
RMS
MEAN
△
▽
ENT
ESC