Chapter
4
4-6.
Zero Calibration
Register the zero point with the sensor unloaded. Also, if there is any initial load, such as a jig, for
measurements, the zero point can be registered with the initial load applied, but the signal input
range needs to be considered as the initial load is subtracted from it.
How to set
1.
Press the [SET.] button. (see P.17)
2.
Press the [Calibration] button. (see P.19)
3.
Press the [Zero Cal.] button.
4.
Press [OK] button after confirming no-load
was applied to the sensor.
4-7.
Increment
(optional if there is no change)
Set the minimum unit of torque such as scale interval and scale division.
≪
Setting value
≫
1, 2, 5, 10, 20, 50, 100 (The display changes by the decimal place.)
How to set
1.
Press the [SET.] button. (see P.17)
2.
Press the [Calibration] button. (see P.19)
3.
Select page 2, and press the [Increment]
button.
SET.
Calibration
Page 1
→
→
SET.
Calibration
Page 2
→
→