SETTINGS AND CALIBRATIONS
40
Chapter
4
SETTINGS
AND CALIBRATIONS
3.
Waveform Reference
Select whether the waveform reference is front or back.
Setting range
Front: The measurement start point becomes the
waveform reference.
Operation
Back:
The measurement stop point becomes the waveform reference.
Main screen
→
Setting
→
First Setting
→
X-axis Setting
→
Wave Reference
4.
Unit
Select a unit.
Operation
Setting range (See "9-3.Unit setting list"on p.149.)
Main screen
→
Setting
→
First Setting
→
X-axis Setting
→
Unit
5.
Zero Calibration
Register the present sensor input as zero.
Press
with the rotary encoder zeroed.
Operation
Setting range (0 to 1000000)
Main screen
→
Setting
→
First Setting
→
X-axis Setting
→
Next
→
Zero Cal.
When the Waveform Reference is set at Back, Change of Section cannot be
made externally.
Point
< Measurement waveform >
Displacement
Load
< Comparison waveform >
Displacement
Load
< In case of “Front” >
Displacement
Load
< In case of “Back” >
Displacement
Load
Front
Back
Front
Back
Same as Back
Same as Front
Measurement waveform
Comparison waveform
0°
0°
Measurement
waveform