5.CALIBRATION
31
5-4-2. Load cell excitation
Select load cell excitation between 10V and 5V.
Refer to the specification of load cell to select suitable excitation please.
Operation
5-4-3. Unit
The unit can select from t, kg, g, N, lb or None unit.
Operation
5-4-4. Decimal place
Decimal place should be selected from 0/0.0/0.00/0.000.
Operation
5-4-5. Capacity
Register the capacity of the scale. The capacity must be within the rated load of load cells combined.
The input range of capacity is between 0 to 99999.
Operation
Notice
Concerning Lock and Setting values to be protected, refer to THE LIST OF INITIAL
SETTING VALUE on Page110.
MODE
CALIBRATION
PAGE
→
→
Press
LOADCELL EXCITATION
→
→
Select 10V/ 5V
OK
→
MODE
CALIBRATION
PAGE
→
→
Press
UNIT DISPLAY
→
Select Unit
→
OK
→
MODE
CALIBRATION
PAGE
→
→
Press
DECIMAL PLACE
→
→
Select 0/ 0.0/ 0.00/ 0.000
OK
→
All of the decimal place in F805AT-CK are fixed except for weight.
* They cannot be changed.
MODE
CALIBRATION
CAPACITY
→
→
Press
Input 0 ~ 99999
→
OK
→