5.CALIBRATION
36
5-6-4. Balance weight
Enter the value of balance weight that is used (actually applied to the load cell) in span
calibration.
Operation
5-6-5. Capacity
Defines the maximum weight that can be applied to the instrument.
Operation
5-6-6. Minimum scale division (weight)
Defines the minimum unit value (resolution) to gross weight, weight density, and load
Operation
ratio measurements.
5-6-7. Excitation voltage
Defines the excitation voltage applied across the bridge circuit of the load cell. For
Operation
better measurements, select a larger value within the allowable range of the load cell.
Input range: 0 – 99999 (5 digits, decimal point is automatically set)
MODE
WEI.CALIBRATION
→
→
BALANCE WEIGHT
Input range: 0 – 99999 (5 digits, decimal point is automatically set
MODE
WEI.CALIBRATION
→
→
CAPACITY
according to the weight)
Input range: 1 – 100 (3 digits, decimal point is automatically set
MODE
WEI.CALIBRATION
→
→
MIN.SCALE DIV.(WEIGHT)
according to the weight)
Allowable value: 5V, 10V
MODE
WEI.CALIBRATION
→
→
LOADCELL EXCITATION