5.CALIBRATION
31
5-6. Weight Calibration
5-6-1. Zero calibration
Zero calibration tunes the instrument’s zero point.
Check the weighing device (load cell) and make sure no load is applied to the sensor.
A foreign matter placed on the load cell or contact to a foreign object can interfere the
calibration.
Make sure that
indicator is off and reading is stabilized. Good calibration
cannot be obtained while readings are fluctuating.
* While the instrument is running,
indicator becomes reversed display (light
blue). The user must either stop the instrument or enter a password (see page 32 for
the password).
Operation
1
)
Press [ZERO CALIBRATION].
2
)
Press
to perform zero tuning and register parameter values.
RUN
RUN
MODE
WEI.CALIBRATION
→
→
ZERO CALIBRATION
OK
Pressing BALANCE WEIGHT,
CAPACITY, or MIN.SCALE
DIVISION key will switch the screen
to the corresponding setting screens.
kg