5.COMPARISON SETTING
21
5. COMPARISON SETTING
■
Comparison mode setting
Select the comparison mode.
Alternatives:
[
HI
-
LO COMP
]
[
DIS.
-
O/U
]
[
DIS.
-
H/L
]
Explanation of each alternative
HI/LO limit comparison (HI
-
LO COMP)
The net or gross weight is compared with the set values.
Final discharge – overweight/underweight comparison (DIS.
-
O/U)
This function is for accurate discharge of the weight value. Overweight/underweight judgment
is made by comparing with the net weight.
Final discharge – HI/LO limit comparison (DIS.
-
H/L)
This function is for accurate discharge of the weight value. Comparisons are made with the net weight.
High/low limit comparisons are made with the net or gross weight.
■
Comparison inhibiting time/ Judging time/ Completion output time
Valid in the discharge mode. Comparisons are not made for a fixed time by this function to prevent
inappropriate operation of control and judgment due to mechanical vibrations related to opening and
closing of valves. The length of the completion signal to be output when weighing is completed can
also be set.
COMP.INH.TIME
Setting range: 0.00 to 9.99 (sec.)
JUDGING TIME
Setting range: 0.00 to 9.99 (sec.)
COMPLETE OUT T. Setting range: 0.00 to 9.99 (sec.)
0
FINAL – SP1
FINAL – SP2
FINAL – CPS
Comparison inhibiting timer
Judging timer