CALIBRATION
31
CALIBRA
T
ION
Chapter
4
4-2.
Calibration Protect
Calibration-related set values can be protected so that they will not be changed by misoperation.
When Cal. Protect is ON, no change can be made while the alarm sounds.
ON
:
Protected
OFF
:
Unprotected
How to set
1.
Press the [SET.] button.
(
refer to page 33
2.
Press the [
Protect / Init.]
button.
3.
Press the [Cal.
Protect]
button.
4.
Select the ON/OFF and determine with
the [OK] button.
4-3.
Calibration Value Selection
(This step may be omitted if there is no change.)
By storing up to four calibration values in the memory, the desired calibration value can be called to
switch the indicated value. Setting values that can be switched are as follows:
■
Calibration value selection by touch panel
Select "0" - "3" by pressing the
button on the CALIBRATION screen.
When not using, set "0".
SET.
Protect / Init.
Page 1
→
→
Calibration Mode Setting
Anarog Input Selection
Zero Calibration
Actual Load Calibration
Equivalent Input Calibration
Increment
Unit
Digital Offset
DZ Limit