CALIBRATION
33
CALIBRA
T
ION
Chapter
4
4.
First select the category, and then
select the unit and determine with
the [OK] button.
Select by scrolling with the
button.
4-6.
Zero Calibration
Set the zero point in no-load condition.
How to set
1.
Press the [SET.] button.
2.
Press the [Calibration] button.
3.
Press the [Zero Cal.] button.
4.
Press [OK] button after confirming no-
load was applied to the sensor.
* Even if the unit is changed, the display
value (calibration value) will not change.
SET.
Calibration
Page 1
→
→