1
EXP. HOLD FUNCTIONS
114
EXP. HOLD FUNCTIONS
Chapter
1
1-5.
Renewal of Hold Value
You can change the operation of renewing the hold value display.
All Time / Hold Stop
Initial value: All Time
How to set
- All Time:
When detection starts, the previous hold value is canceled and the indicated
value is displayed by tracking. When the hold condition is met, the indicated
value is held.
- Hold Stop:
When detection starts, the previous hold value is maintained, and no tracking
state is displayed. If the hold condition is met when the Hold Section ends, the
hold value is renewed. This is, however, effective only when Hold Auto Reset
is ON.
- In the case of Peak hold
SET.
Exp. Hold Set.
Page 1
→
→
Exp. Setting
→
A tracking state is restored by inputting T/H.
Hold Start
Level
Section time
䇭䇭䇭䇭
Section Time
All-Time Renewal
Hold-Stop Renewal
H/E
H/E
H/E
SECTION
SECTION
SECTION
T/H
T/H
T/H
Numerical changes in Hold
Action can be seen because the
hold value is canceled when
detection starts.
Hold value changes can simply
be seen because the hold value
is maintained during detection.
Hold-Stop Renewal & Hold Auto
Reset & Level Trigger
For example, if the Hold Section
is set to "Level + Time," hold
detection is performed with no
external signal, and only the
hold value is displayed.
Summary of Contents for F377A
Page 1: ...18JUL2012REV 3 06 DIGITAL INDICATOR F377A OPERATION MANUAL ...
Page 5: ...What is RoHS 4 M E M O ...
Page 12: ...11 11 STANDARD ...
Page 13: ...12 12 M E M O ...
Page 94: ...93 93 EXPANSION ...
Page 95: ...94 94 M E M O ...
Page 97: ...Expansion Function 96 M E M O ...
Page 152: ......