Motion Detect
Set the parameters to detect stability of calculated values.
If the difference between the current indicated value and the 50-msec-previously indicated value
becomes less than the set count and that condition continues for more than the set time, the
indicated value is regarded as stable.
≪
Setting value
≫
Motion Detect (time)
0.0
~
9.9 [second]
Motion Detect (range)
00
~
99 [count]
CAUTION
●
Zero Tracking is equal to Digital Zero functioning automatically. Therefore, the
Digital Zero limit is also monitored.
●
Zero Tracking does not function if the indicated value has already exceeded the
tracking band.
・
When displacement of the zero point is within the set count of tracking and it
continues more than the set time, it is automatically made zero by Zero
Tracking function.
・
The time (tracking delay) is set in the range of 0.1 - 9.9 sec., and the band
(tracking band) is set in the range of 01 to 99.
If the time is set at 0.0 sec. and the band at 00, the zero tracking function does
not work.
Boundary of zero track
DELAY
Indicated value
+COUNT
-COUNT
BAND
+
0
-
DELAY
Band=count×2
From the point when it returned within the range,
counting will be resumed.
When the time is 0.0 sec. and the width is 00 markings, stable is not detected.