STANDARD INTERFACE
62
STANDARD INTERF
ACE
Chapter
8
8-2.
SI/F
(
Serial interface
)
■
SI/F Indicated value format
For the SI/F, indicated value data exist in two areas: GROSS area and NET area.
Select the data.
《
Setting value
》
0
:
Calculated value/ch1 value
GROSS area
:
calculated value
NET area
:
ch1 value
1
:
Calculated value/ch2 value
GROSS area: calculated value
NET area: ch2 value
2
:
ch1 value/ch2 value
GROSS area: ch1 value
NET area: ch2 value
◇
SI/F Indicated value format setting method
1.
Select setting mode 2.
Press
→
once.
2.
Set "
SI/F Indicated value format
."
Press
twice
⇒
"SI/F function select " is
displayed.
Press
, and the highest digit blinks.
Press
once to make the digit of the SI/F
function select blink.
Input the setting value with
.
(
Initial value
:
0
)
Press
to validate it.
3.
Press
twice to go back to the indicated value
display.
■
SI/F print
《
Setting value
》
0: None
No print command is sent.
1
:
Stable value
When stability of the calculated value is detected, a print command
is sent to the SI/F to make the connected printer print automatically.
When Near Zero is OFF, the calculated value is held until Near Zero
ON after stable ON.
Hold is reset in 1.5 sec. after Near Zero ON.
2
:
Hold value
When hold is reset, a print command is sent to make the connected
printer print the hold value automatically. In this case, the hold value
is forcedly taken in the GROSS area in sending.
Also, not-held calculated values are taken in the NET area.
Ch1 Ch
2
H I
LO
H I
LO
FNC
H I
Ch1 Ch
2
OK
LO
PEAK
HOLD
H I
LO
H I
LO
0
0
0
Undefined
SI/F print
SI/F Indicated value
0: Calculated value
1: Calculated value
2
:
ch1/ch2
《
Setting value
》
/ch1
/ch2
format
ESC