4.CALIBRATION
40
4-3-3. Decimal place
Set the decimal place. The decimal place can be selected from 0.000 / 0.00 / 0.0 / 0.
Setting / 3 : 0.000 2 : 0.00 1 : 0.0 0 : 0
CC-Link
0
46
0 to 3
0
MSB
LSB
RWw0006
RWw0005
RWw0004
Write the code No. and set value.
H
L
Code No.
Set value
remote register
Word size
Console
1) Select code 46.
2) Press the ENT key, and then set the
decimal place.
3) Determine the setting with the ENT key.
Decimal place
Not only weight values but also other register addresses of the F156 are not
given the notion of decimal place. For example, to set the capacity at 100.00
and the minimum scale division at 0.01, set 10000 and 1, and then the decimal
place “2: 0.00,” respectively.