3
Basic operation of the DS252
34
SI/F and indicators
Loadcell indicator
....... From the loadcell indicator connected through the SI/F, indicated value/
net weight/gross weight/tare weight data, etc. are sent at intervals of
approx. 0.3 sec regardless of the value displayed each time. On the DS252
side, recorded data can be selected out of them so that recording can be
performed even if the data is not displayed on the indicator.
Digital indicator
........... From the digital indicator connected through the SI/F, the data of the value
displayed each time is sent at intervals of approx. 0.3 sec.
04, 05: Unit Selection Ach, Bch
Select the unit of recorded data on each channel.
55 types of units are available.
Setting name
04: Unit Selection Ach (DATA UNIT)
05: Unit Selection Bch (DATA UNIT)
Selectable items
See the "List of unit settings" (page 108).
Initial value
02 : kg
Display
Loadcell
DS252
Indicated value
Net weight
Gross weight
Tare weight
SI/F
Gross
Different types of data can be recorded.
Net
indicator
weight
DS252
SI/F
Different types of data cannot be recorded.
Indicated value
Digital
indicator
Gross
weight
Indicated value
0 5 : D A T A U N I T
C h B = 0 2 : k g
0 4 : D A T A U N I T
C h A = 0 2 : k g