2.6  To draw more than 100 cc, (one stroke) free the handle by rotating it 90 degrees and push back 

the piston shaft to its original position without removing the detector tube. Pull the pump handle 
again,  for  the  second  stroke.  These  steps  can  be  repeated  to  get  the  desired  number  of 
strokes.  

  

2.7  The target gas concentration is directly obtained on the detector tube scale.  

 

 

 

 

3. 

INSPECTION OF THE PUMP  

 

 

Before  use,  the  pump  should  be  checked  for  air  leak.  The  following  steps  are 
recommended. 

 

 

 

3.1 Insert a new detector tube without breaking its ends into the rubber tube connector. 

 

 

 

3.2  

Align the red dots and pull the pump handle a full stroke. 

 

 

 

3.3 Wait for three minutes, and release the pump handle by turning 90 degrees on left or 

right and slowly allowing the piston to come back by holding the handle. 

 

 

 

3.4 If it does not come back completely to its original position it means that there is a leak in 

the  pump,  the  cause for  which  could  be  one  of  those  described  in  section  4  under 
maintenance. 

 

 

 

4. 

MAINTENANCE  

  

 

 

The pump needs occasional maintenance.

  

  
  
  
  
  
  
  
  
  
  
  
  
  

4.1 Greasing the barrel and replacing the plunger gasket:  

Summary of Contents for ASP-40

Page 1: ...UNI AIR IN PHOS PRECISION SAMPLING PUMP MODEL ASP 40 STRUCTION MANUAL...

Page 2: ...is a piston and barrel air sampling pump Fig 1 For measurement a detector tube with both ends cut is connected to the rubber tube connector at the end of the barrel When the piston is withdrawn by pu...

Page 3: ...nd reaching 50 cc 1 2 stroke or 100 cc full stroke mark For locking the red dots on the stopper and shaft have to be aligned In this position when the shaft is pulled it gets locked at 50 or 100 cc ma...

Page 4: ...he sample air and produces a coloured stain in the sensing chemical The length of the coloured stain measures the impurity gas concentration in ppm on the tube scale The measurement involves the follo...

Page 5: ...p the end of the tube in the sample gas source while the gas is being drawn through the tube for the allotted sampling time approximately 1 3 minutes as indicated on the tube sheet The sampling is con...

Page 6: ...The following steps are recommended 3 1 Insert a new detector tube without breaking its ends into the rubber tube connector 3 2 Align the red dots and pull the pump handle a full stroke 3 3 Wait for t...

Page 7: ...bber packing of the piston iv Inspect the piston gasket on the end of the plunger for damage To replace the gasket use a small screw driver or other sharp tool to pull off the worn gasket and slip a n...

Page 8: ...d and if necessary replaced or corrected Symptom Corrective action Rubber tube connector is loose Rubber tube connector is worn or damaged Piston sticks or slip too easily One way valve s on the pump...

Reviews: