Maintenance
48
10.2 Maintenance work/Inspection
NOTE
The maintenance work must be carried out in accordance with the inspection and
maintenance schedule! The type and extent of the wear depends on the individual
application and operating conditions. All intervals specified are therefore for
guidance only.
The display shows when the inspection intervals have been reached. Perform and
document inspection, and confirm via the menu that the inspection has been
carried out: MENU
COMMANDS
Check OK
[Enter].
Inspection
Interval
(recommended)
Weekly inspection
Purge gas inlet unobstructed (particularly in
case of frost)
weekly
Exhaust gas line unobstructed (particularly
in case of frost)
weekly
Quarterly inspection
Calibrate device according to manufacturer's
specifications, message in display “Service
[Typ] ZERO”, “Service [Typ] SPAN”,
“Service [Typ] MID”
every 3 months,
latest annually or
when required
Half-yearly inspection
Check integrated filter in the device
every 6 months
Check compressed air supply
1)
(negative
pressure during "drainage")
every 6 months
Check lines for condensate (including all
integrated filters)
every 6 months
Check gas inlets and clean professionally if
necessary
every 6 months
Check fan
every 6 months
Check ejector pump
1)
every 6 months
Check inlet filter (inlet ambient air, filter mat,
ventilator)
every 6 months
Check Peltier cooler
1)
every 6 months
Check fan of Peltier cooler
1)
every 6 months
Yearly inspection
Check air and gas pump
1)
(by performing a
purge gas calibration)
yearly
1)
if installed
Summary of Contents for INCA1021
Page 3: ...3 Dimensions ...
Page 10: ...EU Declaration of Conformity 10 ...
Page 16: ...Safety equipment 16 3 2 Markings and warnings Fig 3 1 Markings and warnings 1 Type plate 1 ...
Page 44: ...Operation 44 ...
Page 46: ...Decommissioning switching off 46 ...
Page 50: ...Maintenance 50 ...
Page 54: ...Service 54 ...
Page 56: ...Associated documents 56 ...
Page 58: ...Disposal 58 ...
Page 60: ...Spare parts 60 ...
Page 61: ...61 16 Annex ...
Page 62: ...Annex 62 EU Declaration of Conformity Flame arrester ...
Page 63: ...Annex 63 ...