Operation
71
Main menu - Options - System - Ignition
NOTE
The combustion calorimeter can only be ignited when the door is closed.
The ignition monitoring is configured in this menu.
The following must be observed for this:
Single ignition or interval ignition must be chosen.
Single ignition:
After device start and the flushing time period (10 s), the ignition starts for the max-
imum set ignition duration. This time can turn out to be shorter if the ignition thresh-
old is reached before the ignition duration elapses. If the ignition threshold is not
reached within the specified time, the combustion calorimeter goes to STOP state.
Interval ignition:
After device start and the flushing time period (10 sec), the ignition starts for a
maximum set ignition duration and is repeated after a pause equal to the ignition
duration. This is repeated until the ignition threshold is reached.
The ignition threshold specifies the differential temperature between the cooling air
and flue gas. Default value is 3 °C.
Main menu - Options - System - Ignition
An update from a memory stick is initiated with this key.
Summary of Contents for CWD2005 SPC
Page 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Page 10: ...Technical data 10 ...
Page 12: ...EU Declaration of Conformity 12 ...
Page 18: ...Safety information 18 ...
Page 22: ...Protective equipment 22 ...
Page 48: ...Description of the work stations operator control elements 48 ...
Page 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Page 87: ...Operation 87 ...
Page 88: ...Operation 88 ...
Page 90: ...Decommissioning Switching off 90 ...
Page 94: ...Maintenance 94 ...
Page 100: ...Troubleshooting 100 ...
Page 102: ...Service 102 ...
Page 104: ...Disposal 104 ...
Page 106: ...Spare parts 106 ...