DT-8806S
10
10. Measuring Operation
10-1. Choosing The Temperature Unit — F1 Function
Press “MODE” button for 2 seconds, the screen displays: F1.
Select ”DOWN” for degrees Celsius, ”UP” for degrees
Fahrenheit.
10-2.Alarme Setup — F2 Menu
Press "MODE” touch for 2 seconds, the screen displays: FI .Press
twice ”MODE” touch to get F2. Select ”UP” to increase the
threshold by 0.1°C (0.1°F), ”DOWN” to reduce it by 0.1°C
(0.1°F).
Note: The alarm threshold default value is 38°C (100.4°F).
10-3. Total Difference — F3 Menu
To adjust the total variation of your the Non-contact Body
Infrared Thermometer.
Press “MODE” button for 2 seconds, the screen displays: F1.
Press ”MODE” button two times to get F3. Select ”UP” to
increase the difference by 0.1°C (0.1°F), “DOWN to reduce itby
0.1°C (0.1°F).In the cases of seasonal or environmental
changes a verification and adjustment should be carried out.
Note: This function is only effective Body
10-4. Buzzer ON/OFF — F4(F3) Menu
Press ”MODE” button for 2 seconds, the screen displays: FI.
Press three(two) times "MODE" button to get F4(F3).